Please contact customer service
Maximum temperature 1200 ℃, 16L Furnace volume, Heating time ≤ 30 minutes. Vacuum forming polycrystalline mullite fiber Furnace insulation, special alloy heating wire three-sided heating, Temperature uniformity is good. With multiple Safety protection such as power failure and over-rise alarm.
Category : Muffle Furnace
Brand : TaiSiTe
SKU : NB029207
Service :The delivery and service By NBCLab.Com
TAISITE MFLC-16/12P ceramic muffle furnace is a laboratory heating equipment with a maximum temperature of 1200 ° C. It adopts an integrated lightweight and compact design, furnace volume is 16L, equipped with imported program temperature control digital display instrument, controlled by microcomputer PID, temperature control accuracy can reach +/- 1 ° C.
1. Integrated design, lightweight and compact structure. 2. Vacuum forming polycrystalline mullite fiber furnace effective insulation, special alloy heating wire three heating provides good Temperature uniformity. 3. Open the door and power off, and the reset is delayed to ensure safety. 4. Double shell air insulation, low surface temperature. 5. Exhaust device, conducive to ashing. 6. High Accuracy Microcomputer Temperature Controller and Precision Sensor are well combined for accurate temperature. 7. Dedicated function keys to achieve temperature setting. 8. Auxiliary menu, realize over-rise alarm, deviation correction, menu lock, over-rise preventer. 9. Parallel side-opening furnace doors to keep the hot end away from operation.
The muffle furnace is heated on three sides by a nickel-chromium-aluminum alloy heating element, and the heat is effectively kept warm by a vacuum-formed polycrystalline mullite fiber furnace. The temperature control is coordinated by a microcomputer PID controller and a K-type Thermocouple Sensor to achieve accurate temperature setting and maintenance.
Material ashing, sample heat treatment, laboratory high temperature experiment
The information does not mention
Connect the power supply (AC220V/27.3A 50/60Hz). 2. Use the touch of five buttons to set parameters such as temperature and time (the P-type instrument supports 4 Linear dispersion 40-step programs to run). 3. Start the equipment to start heating, Heating time ≤ 30 minutes to 1200 ℃. 4. The timing function can be used during operation (0 to 999.9 hours). 5. After the experiment, wait for the furnace body to chill down and open the door to power off (the equipment has a door control function).
* The equipment has the functions of door opening and power failure and reset delay. Please pay attention to safety during operation. * The surface temperature of the furnace body may be high, please take care to avoid burns during operation. Do not exceed the maximum temperature of the device to 1200 degrees Celsius. The equipment is equipped with an overcurrent and leakage protection switch to ensure power safety. * When using the exhaust device, please ensure that it is well ventilated. * Device weight (gross weight 86kg) when handling.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.