Please contact customer service
Maximum temperature 1200 ℃, 36L large volume Furnace, using vacuum forming polycrystalline mullite fiber insulation, three-sided heating to ensure Temperature uniformity, with program temperature control, open the door power, over-rise alarm and other multiple safety functions.
Category : Muffle Furnace
Brand : TaiSiTe
SKU : NB029209
Service :The delivery and service By NBCLab.Com
TAISITE MFLC-36/12P ceramic muffle furnace is a laboratory high temperature equipment with a maximum temperature of 1200 ° C. It adopts an integrated lightweight and compact design, the furnace volume is 36L, equipped with a High Accuracy microcomputer PID temperature controller and K-type Thermocouple, the temperature control accuracy can reach +/- 1 ° C. It also has various functions such as program operation and timing.
1. Integrated design, lightweight and compact structure. 2. Vacuum forming polycrystalline mullite fiber furnace, good thermal insulation effect; special nickel-chromium-aluminum alloy heating wire three-sided heating, Temperature uniformity. 3. Double shell air insulation, low surface temperature. 4. Parallel side open furnace door design keeps the hot end away from the operator. 5. Equipped with exhaust holes, which is conducive to ashing and other processes. 6. High Accuracy Microcomputer Thermostat Combined with Precision Sensor, Temperature Accuracy, Resolution 1 ℃. 7. Dedicated function keys enable temperature setting, making operation convenient. 8. Auxiliary menu provides over-rise alarm, deviation correction, menu lock, over-rise prevention and other functions. 9. Multiple safety designs: over-lift alarm, menu lock, over-lift prevention, door opening and power failure, restart delay, over-current and leakage protection switch.
The muffle furnace precisely adjusts the Power supplied to the special alloy heating wire in the furnace through a microcomputer PID Control system. The heating wire adopts a three-sided Heating mode, which converts electrical energy into heat energy, so that the temperature in the vacuum forming polycrystalline mullite fiber furnace rises uniformly and stabilizes at the set value. K-type Thermocouple monitors the temperature in the furnace in real time and feeds it back to the temperature controller to form a closed-loop control, thereby achieving High Accuracy Temperature Control.
Material sintering, ashing experiment, heat treatment, ceramic processing, laboratory high temperature treatment
The data did not mention specific compliance criteria.
1. Connect the AC380V power supply. 2. Set the target temperature, program or timing parameters by tapping the button or Touchscreen of the thermostat. 3. Start heating, the furnace body starts to heat up. 4. During operation, the monitoring status can be displayed through the observation window or instrument. 5. After the program runs or reaches the scheduled time, the equipment stops heating. 6. After the furnace temperature is naturally chilled down to a safe range, the furnace door can be opened to remove the sample.
Equipment needs to be installed and operated by professionals to ensure good grounding. • When powered on and running at high temperature, do not touch the furnace body to prevent burns. • When opening the door, the device will automatically power off. There will be a delay in repowering after closing the furnace door. Please do not force the operation. • Flammable, explosive, volatile and corrosion items are strictly prohibited in the furnace. • Sample placement should not block the vent hole and ensure that it is in the effective heating zone of the furnace. • When heating up and high temperature operation, the furnace body should be well ventilated. • Clean the furnace in time after use, but avoid moisture in the furnace material. Regularly check whether the electrical wiring and safety devices are in good condition. • Before moving or servicing, be sure to cut off the power and make sure the furnace is completely chilled down.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.