Overview
The ANGNI AD18 high-speed dispersion Mulser is driven by a miniature high-power high-speed motor imported from Germany, which is suitable for stirring and mixing low to medium and high viscosity liquids in the laboratory, dispersing homogeneous granular materials, tissue cell crushing, etc. in the liquid medium. The machine is composed of a high-speed motor, a dispersed homogeneous working head and a machine base. The driving host adopts an engineering plastic and a die-casting aluminum alloy body for stable operation. The working head is refined from stainless steel and has a wide range of applications. Rotation speed can be adjusted to ensure effective and controllable experimental operation.
Features
1. Driven by miniature high-power high-speed motors imported from Germany
2. The drive main engine adopts engineering plastic and die-casting aluminum alloy body, and the running state is stable
3. The working head is made of good stainless steel and has a wide range of experimental applications
4. Rotation speed adjustable control to meet the experimental Rotation speed correction at any time
5. With restart protection function
6. Adaptable to a variety of working heads (8G, 10G, 12G, 18G, 24G) optional
Principle
Driven by a high-speed motor dispersed homogeneous head in the liquid medium high-speed swirl/spin, so that the experimental medium in the material liquid through extrusion, strong Impact and depressurized expansion of the triple role, to achieve material refinement, uniform mixing, to prevent or reduce the material liquid delamination.
Applications
Tissue dispersion, sample preparation, enzyme processing in the food industry, industrial, cosmetic industry, paints industry and Petrochemicals in the field of biotechnology
Standards
Rated voltage AC220V, Rated Frequency 50/60HZ, Duty Cycle S1 (continuous)
Steps
1. Select the appropriate working head specification
2. Rotation speed setting
3. Immerse the working head into the liquid medium (minimum immersion height 15mm, maximum immersion height 105mm)
4. Start the device for decentralized emulsification
5. Rotation speed can be adjusted at any time according to experimental needs
Notice
• Suitable for liquid material experiments with viscosity below 0.2Pa.s and temperature below 80 ° C
• Maximum processing viscosity is 5000mPa.s
• Maximum operating temperature is 120 ° C
• Permissible Ambient Temperature ≤ 40 ℃
• Allowable relative humidity ≤ 80%
• The height of the working head immersed in the liquid should be within the range of 15-105mm
• Pay attention to the suitability of liquid-solid and liquid-liquid materials