Overview
The ANGNI AD18 high-speed dispersion Mulser is driven by a miniature high-power high-speed motor imported from Germany, which is suitable for stirring and mixing low to medium and high viscosity liquids in the laboratory, dispersing homogeneous granular materials, tissue cell crushing, etc. in the liquid medium. The machine is composed of a high-speed motor, a dispersed homogeneous working head and a machine base. The driving host adopts an engineering plastic and a die-casting aluminum alloy body for stable operation. The working head is refined from stainless steel, which has a wide range of experimental applications. Rotation speed can be adjusted to ensure effective and controllable experimental operation.
Features
1. Adopt miniature high-power high-speed motors imported from Germany
2. The drive main unit is made of engineering plastic and die-casting aluminum alloy, providing stable operation
3. The working head is made of stainless steel and has a wide range of experimental applications
4. Rotation speed adjustable control to ensure effective and controllable experimental operation
5. With restart protection function
6. Adapt to a variety of working head specifications (8G, 10G, 12G, 18G, 24G)
Principle
The high-speed motor drives the dispersed homogeneous working head to swirl/spin in the liquid medium, so that the experimental medium is squeezed in the material liquid, the triple effect of strong Impact and depressurized expansion, the material is refined and uniformly mixed, and the delamination phenomenon is prevented or reduced.
Applications
Tissue dispersion, sample preparation, enzyme processing in the food industry, industrial, cosmetic industry, paints industry and Petrochemicals in the field of biotechnology
Notice
* Suitable for liquid materials with a viscosity below 0.2Pa.s
The liquid temperature must be lower than 80 ° C
* Maximum processing viscosity does not exceed 5000cp
Maximum operating temperature does not exceed 120 ° C
* Permissible Ambient Temperature not exceeding 40 ℃
Permissible relative humidity not exceeding 80%
* Minimum immersion liquid height is 50mm
* Maximum immersion liquid height is 165mm