Overview
The FANGRUI LV-SSR digital rotational viscometer adopts a coaxial cylindrical structure and is equipped with a small sample adapter. Only 7ml-13ml of samples are required for each measurement (21ml for No. 0 rotor). Optional with circulating thermostatic water jacket with Thermostatic bath temperature control. Easy to operate, accurate measurement, especially suitable for viscosity detection of precious and small samples.
Features
1. The front design of the horizontal bubble, the level can be adjusted by sitting, and the operation is convenient
2. With timing measurement function
3. Adopt coaxial cylindrical structure
4. Measurement accuracy to two decimal places
5. Easy and reliable operation
6. Use a small number of sample adapters, and the amount of sample required for testing is minimal
7. Equipped with two RS232 interfaces, it can directly connect to micro thermal printers and computers
Principle
The coaxial cylinder structure is used to measure the fluid viscosity through the resistance generated by the rotor swirling/spinning in the sample, and the small sample adapter is used to achieve micro sample detection.
Applications
Oil, paints, coatings, Stock, Spinning & weaving, Food, Adhesive, Cosmetics and other fluid viscosity detection and control
Standards
Measurement error +/- 2% (Newton liquid), repeability error +/- 1% (Newton liquid)
Steps
Step 1 Adjust the instrument level
2. Choose the right rotor and Rotation speed
3. Add an appropriate amount of sample
4. Initiate measurement
5. Read and record data
6. Can be connected to a printer or computer through RS232 interface to output data
Notice
• Ensure the level of the instrument before measurement
• Choose the appropriate rotor and Rotation speed according to the sample viscosity range
• Strictly control the amount of samples added
Maintain a stable measurement environment
• Regularly clean the rotor and adapter
• Make sure the interfaces match when connecting external devices