Overview
XINGYUN XY-100MW-T Moisture Analyzer is a fully automatic heating moisture analysis instrument, using halogen lamp heating source and imported HBM weighing Sensor, Capacity value of 110g, moisture measurement range of 0.00~ 100.00%, suitable for all kinds of samples Moisture content accurate determination.
Features
1. The dual-drive ultra-quiet opening of the fully automatic heating cabin door prevents the impact of artificial flip-off on weighing and meets the requirements of automatic testing
2, 5 inch high definition intelligent touch display, Linear dispersion real-time dynamic analysis, direct display moisture percentage/dry weight percentage/dry weight weight
3. 230 ° C limit heating temperature, meeting the vast majority of sample tests
4. The sample is effectively placed in the heating cabin with a height of 23cm.
5. Sample testing can be viewed in a window to monitor sample test changes in real time
6. 20 sets of effective storage addresses to meet the test settings for different samples
7. 20 sets of recent test records, ready to retrieve and print test results
8. It can meet the maximum weighing capacity of 110g
9. Imported HBM Weighing Sensor
10. Four layers of stainless steel heat the inner compartment to improve the efficiency of sample testing
Principle
The instrument heats the sample through the halogen lamp heating source, uses the PT100 temperature sensor to precisely control the heating Temperature range from RT + 40 to 199 ℃, and monitors the sample weight change in real time through the imported HBM weighing Sensor, calculates the moisture percentage and dry weight data, and finally displays the analysis results on the touch screen.
Applications
Food, chemical, pharmaceutical, agriculture, environmental protection, materials
Standards
GB/T 29249-2012、JJG 658-2010
Steps
1. Turn on the power supply and turn on the instrument
2. Select or set test parameters through the touch screen
3. Place the sample on the pan
4. Close the heating compartment door and start the automatic test program
5. Linear dispersion and data display
6. View and save the results after the test is completed
7. You can choose to print or retrieve history
Notice
• Operating Environment Temperature should be kept within 5~ 35 ℃
• Avoid use in humid or vibrating environments
• Capacity should not exceed 110g
• Regularly clean heating tanks and pans
• Avoid direct contact with high-temperature components during operation
• Turn off the power in time after use