Please contact customer service
Measurement range 0-1250μm, resolution 0.1μm. Support one/two point calibrating, Data storage, statistical analysis and USB communication function.
Category : Coating Thickness Gauge
Brand : PoRuiKe
SKU : NB029999
Service :The delivery and service By NBCLab.Com
PORUIKE BCT-210C (F1) is a split High Accuracy Coating thickness gauge. It adopts magnetic Induction and eddy current thickness measurement methods. It is suitable for measurement platings, paints layers, enamel layers, aluminum tiles, copper tiles, Babbitt tiles, phosphating layers, paper and other Coating thickness measurement. It can also be used for thickness measurement of hull paints and underwater structural attachments. This instrument is widely used in manufacturing, metal processing, chemical industry, commodity inspection and other testing fields.
1. Full Chinese interface operation, more convenient. 2. Adopt two thickness measurement methods of magnetic and eddy current. 3. Can be equipped with a variety of probes (such as F400, F1, F1/90 °, F10, CN02, N1, N10, F5, etc.). 4. Measured value outside the limit can be automatically alarmed, and the batch Measured value can be analyzed by histogram. 5. LCD liquid crystal display, backlight control can be switched on and off. 6. With menu operation function. 7. With the function of communicating with the PC, Measured values and statistical values can be transferred to the PC for further processing. 8. With error prompt function, error prompt is given through screen display or beep. 9. There are two shutdown methods: manual shutdown and automatic shutdown. 10. It is designed with 2 AA batteries, which is convenient and simple to replace. 11. Adopting an international 9-pin socket and a standard USB interface, it can communicate with the host computer through software.
This instrument adopts two thickness measurement methods: magnetic Induction and eddy current. The magnetic Induction method is used to measure the nonmagnetic Coating thickness on the magnetic substrate, and the eddy current method is used to measure the insulation Coating thickness on the nonmagnetic metal substrate. By contacting the coating surface with the Probe, the instrument measures and calculates the thickness of the coating according to the principle of electromagnetic induction.
Manufacturing, metal processing, chemical industry, commodity inspection
1. Power on and select the system language (Chinese). 2. Select the appropriate Probe according to the measurement requirements and connect it. 3. Select the calibrate mode (one point calibrate or two point calibrate) and calibrate. 4. Measurement parameters, such as limits, units (μm), etc. 5. Select Measurement mode (Continuous Measurement mode or Single Measurement mode). 6. Place Probe vertically on the surface of the coating to be tested for measurement. 7. Read the measurement results on the LCD display. 8. The measurement data can be transmitted to the PC through the Interface. 9. After use, you can choose to shut down manually or wait for automatic shutdown.
* Ensure that the Probe is in vertical contact with the coated surface for accurate measurement. * Probe and calibrate modes are selected according to the substrate material and coating type before measurement. * Pay attention to the working Ambient Temperature (0-40 ℃) and relative humidity (20~ 90% RH) of the instrument. * For the F1 Probe, the minimum measurement surface diameter is φ 7mm, the minimum radius of curvature is convex 1.5mm, and the thinnest substrate is 0.5mm. Regularly check the battery level to ensure that the instrument is working properly. * Avoid using the instrument in a strong electromagnetic interference environment. * The instrument has an error prompt function. If there is an error, please check it according to the prompt. * When storing measurement data, note that the maximum Storage capacity of the instrument is 500 Measured values.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.