Overview
BOLEL PURIFICATION EQUIPMENT VD-850 (Remote Control) Desktop Laminar Flow Cabinet is a highly versatile local purification device with a vertical wind design, suitable for electronics, security, precision instruments, instrumentation and other industries. The device is compact and portable, and can be placed on a general workbench for operation, making it easy to use in small studios.
Features
1. Vertical manifold, quasi-closed Glass damper, easy to operate.
2. Adopt an adjustable air volume fan system to ensure that the working wind speed is always in a satisfactory state.
3. Compact, can be freely placed on a general workbench for operation, making it convenient for small studios to use.
4. Use remote infrared control.
The experimental area is made of 304 stainless steel to provide convenience for the operation certificate.
Principle
The workbench adopts the principle of vertical laminar flow, and provides clean air through high-efficiency filters to form vertical airflow to ensure that the working area reaches the 100-level clean level (US LIANBANG standard 209E). The adjustable air volume fan system maintains a stable wind speed to ensure the purification effect.
Applications
Electronics, security, precision instruments, meters
Standards
American LIANBANG Standard 209E
Steps
1. Turn on the power supply (220V 50HZ).
2. Use the remote infrared remote control to turn on the device.
3. Adjust the air volume to the wind speed range of 0.3 to 0.6m/s as required.
4. Open the quasi-closed Glass throttle and start the operation.
5. After the operation is completed, turn off the device and disconnect the power supply.
Notice
• Power supply voltage is 220V 50HZ to avoid voltage instability.
• Keep the work area clean during operation to avoid contamination.
• Regularly check the status of the filter and replace it in time to ensure the purification efficiency.
• The noise of the equipment should not exceed 62dB (A) during operation, and if there is any abnormality, it should be repaired in time.
• Avoid use in vibration environment, vibration half peak should be ≤ 3μm.
• Turn off lighting and sterilization lamps after use to save energy.