Overview
The LEICI PCN-1-01 Cyanide Ion Electrode is an Electrochemical sensor for the measurement of cyanide Ion concentration. Measurement range is 10 ² to 10 mol/L. The Electrode features an Organic Glass case with a BNC (Q9) connector for Ion concentration detection in laboratory and industrial environments.
Features
1. Measurement range, covering 10 ² to 10 mol/L Cyanide Ion concentration
2. Sensitive film adopts salt film design, Response is sensitive
3. Shell material is Organic Glass, corrosion resistant and lightweight
4. Operating temperature range, 5-60 ℃
5. Standard BNC (Q9) connector, strong compatibility
Principle
Cyanide ion Electrode works based on the principle of ion selective Electrode, and its sensitive membrane has a specific Response to cyanide ions. When Electrode is immersed in the solution to be tested, the membrane potential changes with the cyanide Ion concentration. The potential difference is measured by the matching pH Meter or ion meter, and converted into the Ion concentration value output.
Applications
Water quality monitoring, environment testing, industrial process control, laboratory analysis
Standards
BNC (Q9) Interface standard
Steps
1. Connect the Electrode to the matching pH Meter or ion meter via the BNC interface
2. Electrode calibrated with standard solution
3. Electrode immersion in the sample solution to be tested
4. Record the Measured value after waiting for the reading to stabilize
5. Clean Electrode After Use and Store Properly
Notice
• Electrode sensitive film to avoid mechanical damage
• Electrode cleaning with deionized water before and after use
• Electrode head should be kept moist during storage
• Operating temperature should be controlled within the range of 5-60 ° C
• Electrode performance is regularly checked with standard solution