Overview
LEICI PCU-1-01 Copper Ion Electrode is an ion electrode for the measurement of copper Ion concentration in solution. Measurement range is 10-1 to 5 × 10-7 mol/L, suitable for use with pH Meters or ion meters.
Features
1. Measurement range: 10-1 to 5 × 10-7 mol/L
2. Sensitive film adopts salt film design
3. Shell material is Organic Glass
4. Dimensions: φ 12 × 120mm
5. Operating temperature range: 5-60 ℃
6. Adopt BNC (Q9 type) standard connector
Principle
The Electrode is based on the principle of ion selective Electrode, and generates a potential signal related to Ion concentration through the selective Response of copper ions to the sensitive film, which is measured and converted into a concentration value by supporting instruments.
Applications
Water quality monitoring, environment testing, industrial process control, laboratory analysis
Standards
BNC (Q9) Interface standard
Steps
1. Connect Electrode to pH Meter or ion meter via BNC interface
2. Electrode immersion in the solution to be tested
3. Record the Measured value after waiting for the reading to stabilize
4. Clean Electrode After Use and Store Properly
Notice
• Electrode is nondestructive before use
• Keep the solution temperature in the range of 5-60 ° C during measurement
• Electrode sensitive film to avoid mechanical damage
• Electrode should be cleaned promptly after use