Overview
AISRY ASR-1021 (5KG) touch screen single column tensile strength testing machine is a kind of equipment suitable for mechanical performance testing of various materials. It is mainly used for tensile strength, peel strength, deformation rate, breaking force, disbonding force, opening force and other performance tests of Plastic film, composite film, Adhesive, adhesive tape, medical patch, protective film, release paper, rubber, artificial leather, paper fiber and other products.
Features
1. Adopt touch screen operation interface, intuitive and convenient.
2. The force value resolution is high, reaching 1/200,000.
3. The test speed range is wide, ranging from 0 to 300mm/min.
4. Provide a variety of force values and stress units to choose from.
5. With upper and lower limit safety settings and test breakpoint sensing and other shutdown methods.
6. Equipped with a micro printer, it can directly output test results.
7. Adopt a variable-speed motor as the power system.
Principle
The testing machine is driven by a variable speed motor to apply tensile or disbonding force to the sample. The force value Sensor detects the force condition in real time and transmits the data to the control system. The system controls the testing process according to the set parameters (such as speed, stop conditions), and calculates and displays the tensile strength, Peel Strength, deformation rate and other related performance indicators.
Applications
Adhesive Tape, Automotive, Pottery and porcelain, Composites, Construction, Food Equipment, Metal, Paper, Encasement, Rubber, Spinning & weaving, Wood, Communication and Various Shaped Materials
Steps
Connect the power supply (1PH, AC220V, 50/60Hz).
2. Install the appropriate tensile fixture according to the test requirements.
3. Set test parameters on the touch screen, such as test speed, force unit, etc.
4. Clamp the specimen correctly on the fixture.
5. Start the test and the device will run automatically.
6. After the test is completed, the results can be output through the micro-printer.
7. Turn off the power and clean the equipment.
Notice
* Power supply voltage meets requirements (1PH, AC220V, 50/60Hz).
Please confirm the upper and lower limit safety settings before operation to ensure test safety.
Select the appropriate fixture according to the sample type and install it correctly.
* The weight of the equipment is about 40kg, so pay attention to safety when moving.
* Regular maintenance to ensure Measurement accuracy ( +/- 1%).