Overview
The LABTOP IC170 is a general-purpose chill down circulating water machine, which belongs to the IC series. This series of products is widely used, and the minicomputers can be used to support chill down laboratory equipment. The IC170 is an Air-Cooled all-in-one machine with fan cooling, multiple protection functions, and provides alarm output for external master devices to ensure safety. The device supports multiple Communication interfaces and analog/switch input and output, which can realize remote control and status query.
Features
1. LCD screen, displaying the actual temperature, set temperature and return difference at the same time, the operation is intuitive, and the keyboard is automatically locked when the button operation expires.
2. A variety of pressure pumps are available, and the output pressure of the pump is adjustable to meet the application requirements of different pressures and Flow rates.
3. Equipped with standby antifreeze function of water pump.
4. Support remote control of the switch machine.
5, using brand Compressor, silent fan and silent pump, quiet work, suitable for quiet environment.
6, refrigeration system with high voltage protection and overcurrent protection function, effectively prevent Compressor overload.
7. When alarming, the light and sound are reminded at the same time, and the alarm code is directly displayed, which is convenient for quickly determining the cause of the failure and the repair method.
8. Adopt caster design to facilitate machine movement.
Principle
The chill down circulating water machine drives the refrigeration cycle through the Compressor, and absorbs the heat generated by the equipment through the refrigerant and transfers it to the heat dissipation end. As an Air-Cooled all-in-one machine, it uses a fan to dissipate heat into the air, so as to cool down the chill down liquid. The cooled down liquid is transported to the external equipment that needs to be chilled down through the circulation pump, absorbs heat and then returns to the water machine for chill down, and so on to maintain the set temperature.
Applications
Universal chill down circulating water machine Widely used, small universal chill down circulating water machine can be used for laboratory equipment matching chill down
Steps
1. Place the device in a stable position with Permissible Ambient Temperature (5~ 32 ° C) and relative humidity (≤ 80%), and connect the power supply (200~ 230VAC/50Hz).
2. Connect the chilled down pipeline to the external device to be chilled down through the device interface (4 points inner teeth).
3. Set the required temperature (range 5 to 35 ° C) and Miscellaneous parameters on the LCD screen.
4. Start the device and observe the actual temperature, setting temperature, and operating status on the display.
5. During the operation of the device, remote control or status query can be performed through supported Communication interfaces (such as RS232/RS485/USB).
6. After use, shut down according to the procedures.
Notice
• Please ensure that the device operates at the allowable Ambient Temperature (5~ 32 ° C) and relative humidity (≤ 80%).
• Power supply voltage 200~ 230VAC/50Hz, and properly grounded.
• When the equipment is running, do not block the fan from entering and leaving the trend to ensure good heat dissipation.
• Pay attention to the alarm prompt, and check and deal with the fault in time according to the displayed alarm code.
• When moving the device, please use the casters to move smoothly to avoid violent vibration.
• Regularly check the chilled down level and piping connection to ensure no leakage.
• Non-professionals should not disassemble or repair the internal parts of the equipment.