Overview
HZLB LRG-80Y Intelligent Artificial climate box is a precision environment simulation device with a volume of 80L. It is controlled by a large-screen LCD program, which is intuitive and convenient to operate. It can precisely control temperature, Humidity and light, providing an ideal artificial climate environment for plant growth, seed germination, etc. The product is made of fully liner stainless steel and can be equipped with a variety of intelligent functions.
Features
1. Large-screen LCD program control, Chinese-guided operation, blue backlight, easy to view at night.
2. Intuitively display the running process, light level, setting and box temperature Humidity, Remaining time, Ambient Temperature and protection temperature.
3. 1-99 time periods can be set, with a time range of 0-999 hours for each time period, to achieve automatic program conversion.
4. Program control Illumrination degree 0-3 adjustable (strong light 0-8 adjustable).
5. The liner protection temperature can be set, and the heating power supply can be automatically cut off when it exceeds the temperature to protect the sample.
6. With Ultra Temperature Sensor Abnormal Protection Function to Ensure Instrument and Sample Safety.
7. Optional with full Light spectrum plant growth lamp (disinfection mold lamp), which is conducive to plant growth.
8. Using LED Cold Illuminant, the beam of light does not generate heat, extracting useful light that is conducive to plant growth.
9. With power-down memory and power-down time automatic compensation function, the original working state can be continued after power failure.
10. With Sensor (Ambient Temperature, Protection Temperature, Cabinet Temperature) fault identification function.
11. Optional sterilization function.
12. Optional network data extraction function, transfer data to the cloud platform through GPRS, support computer remote viewing, downloading and analysis of data Linear dispersion.
13. Optionally view data and Linear dispersion through the WeChat Mini Program on the mobile terminal.
Principle
The climate box is controlled by a microcomputer program Control system. Temperature control is achieved by the heating system and the Compressor refrigeration system working together; Humidity control is adjusted by the humidification system; and light is provided by the programmed LED Illuminant. The system continuously monitors the environment parameters in the box, compares them with the set values, and adjusts them dynamically through the actuator to maintain the set stable climate conditions. At the same time, the system has multiple Safety protection mechanisms.
Applications
Plant growth research, seed germination testing, microbial fostering, insect feeding, material aging testing, biological product preservation
Steps
1. Place the equipment in a working environment that meets the requirements (temperature 0-35 ° C, no corrosion gas), connect the power supply (220V +/- 22V, 50Hz +/- 0.5Hz).
2. Turn on the power switch, and set the parameters according to the Chinese operation instructions through the large-screen LCD interface.
3. Set the required operating procedures, including temperature, Humidity, light level for 1-99 periods, and running time for each period (0-999 hours).
4. Set safety parameters such as liner protection temperature as required.
5. After confirming that the parameters are set correctly, start the device to run.
6. During operation, various real-time data and operating status can be visually viewed through the screen.
7. After the operation is completed, or when an interruption is required, shut down the device according to the procedures.
Notice
* Please ensure that the equipment operates at the specified Ambient Temperature (0-35 ° C) and without corrosion gases.
* Power supply voltage must meet the requirements of 220V +/- 22V, 50Hz +/- 0.5Hz.
* The Compressor has a delay protection time of 3 minutes, and frequent switching may affect the life of the equipment.
Pay attention to regularly check and replenish the water in the humidifier water tank (capacity 5L).
* When the equipment is running, do not open the door frequently to avoid affecting the environment Stability in the box.
* If the network or mobile phone monitoring function is optional, please ensure that the device is in the network signal coverage area.
Before cleaning or maintenance, be sure to disconnect the power supply of the equipment.