Overview
The HZLB LDS-1G Grain Moisture Analyzer is a precision instrument for the rapid determination of moisture in various grains and Non-metallic granular substances. It features automatic measurement, temperature compensation and user-defined calibrating for a variety of agricultural and processing scenarios.
Features
1. Automatic weighing, automatic temperature compensation, automatic moisture measurement, automatic shutdown.
2. Equipped with empty machine moisture correction function and new compensation mode.
3. AC/DC dual-use, Flexible Power Supply.
4. Users can calibrate and correct errors by themselves.
5. Equipped with the function of calculating the average value and the function of automatic capacity conversion display (unique to the LDS-1G model).
Principle
The instrument automatically calculates the Moisture content of the sample by measuring changes in physical properties such as the dielectric constant of a Basis Weight sample (usually 150 grams), combined with pre-set or user-calibrated calibrated Linear dispersion. The measurement process integrates automatic weighing and temperature compensation to improve the accuracy of the results.
Applications
Determination of Moisture of Non-metallic Granular Substances such as Cereals, Wheat, Rapeseed, Soybean, Vegetable Seed, Corn and Feed
Standards
The information does not mention
Steps
1. Boot preparation: Load four AA dry batteries or connect an external 9V DC regulated power supply.
2. Sample sampling: Take a Basis Weight sample (usually 150 grams, except for special samples) and put it into a measurement container.
3. Start measurement: The instrument automatically performs weighing, temperature compensation and moisture measurement, and the process is less than or equal to 10 seconds.
4. Reading result: Measured value of moisture is read on the display.
5. Follow-up operation: You can use the function of calculating the average value, or perform user-generated calibration and error correction as required.
6. After use: The instrument has an automatic shutdown function, or manual shutdown.
Notice
• Measured objects are limited to Non-metallic granular substances such as cereals, wheat, rapeseed, soybean, vegetable seeds, corn, feed, etc.
• Please make sure to use the instrument at an Ambient Temperature of 0~ 40 ° C.
• Accuracy ≤ +/- 0.5% (main moisture range), Repeatability ≤ 0.2%, Measurement range is 3~ 35%.
• The sampling amount is usually 150 grams of Basis Weight, except for special samples.
• The instrument has pre-calibrated 20 common varieties, which can be directly measured; Miscellaneous varieties require users to calibrate at three points.
• Please use the designated AA dry battery or an external 9V DC regulated power supply.