Overview
PINGXUAN DC1006-II High and Low Temperature Thermostatic bath is a device that provides an accurate and stable temperature environment for scientific research and production. It is widely used in many fields to provide users with a cold and hot controlled, uniform and constant temperature field source when working. It can be used for constant temperature experiments or tests on experimental samples or products, and can also be used as a direct or auxiliary heating/cooling heat source or cold source.
Features
1. Adopt self-developed special software temperature control technology, equipped with High Accuracy PT100 Sensor and imported electronic components, accurate temperature control.
2. The software system supports temperature correction, with resolution up to 0.01 ° C to ensure accurate display temperature.
3. With over-temperature protection and alarm function, the alarm temperature can be set, and the load can be automatically cut off when overheating, which is safe and reliable.
4. Intelligent PID Control System Temperature Stability, Automatic Adjustment Parameters, Also Support Manual Adjustment by Special Users.
5. The software digital locking function can prevent misoperation and ensure that the experimental data is correct.
6. Accurate temperature control allows for rapid and stable temperature in the working tank.
7. Adopt high-efficiency fully enclosed Compressor for cooling, with fast cooling speed and multiple protections such as overheating and overcurrent.
8. Equipped with professional domestic circulation and external circulation system, domestic circulation ensures uniform temperature, external circulation can establish a second thermostatic field outside the machine or as a cold source.
Principle
The device monitors the medium temperature in the working tank in real time through a High Accuracy temperature sensor (PT100), and feeds the signal back to an intelligent control system developed based on special software. The system precisely adjusts the power output of the heating or cooling system (fully enclosed Compressor) through the PID algorithm to achieve fast and stable temperature control. At the same time, its domestic circulation system continuously Stirs the medium in the tank to ensure uniform temperature; the external circulation pump can pump out the thermostatic liquid to establish an external thermostatic environment or chill down the external experimental vessel.
Applications
Research departments of petroleum, chemical industry, electronic instrumentation, physics, chemistry, biological engineering, life science, light industry food property testing, chemical analysis, etc., colleges and universities, enterprise Quality control and production departments
Standards
The data did not mention specific compliance criteria
Steps
1. Turn on the power supply and turn on the device.
2. Set the required target temperature and over-temperature alarm temperature through the control system.
3. The equipment starts to work, the cooling or heating system starts-up, and the domestic circulation system operates to make the temperature uniform.
4. After the temperature reaches the set value and stabilizes, the experiment or test can be carried out.
5. If you need to use the external circulation function, connect the external pipeline.
6. After the experiment is over, press the "Power" button to turn off all functions, and the device will still display the natural temperature change process in the tank.
Notice
* Please ensure that the equipment is placed in a smooth and well-ventilated environment before use.
* When setting the temperature, do not exceed the rated Temperature range of the device (-10~ 160 ℃).
Do not allow unrelated personnel to operate the device. You can use the software lock function to prevent misoperation.
* Pay attention to the equipment alarm prompt, please check and deal with the over-temperature alarm in time.
* When using the external circulation function, please ensure that the external pipeline connection is correct and leak-free.
After the equipment is turned off, the temperature in the tank will naturally rise or fall. Please pay attention to observation or handling.
Regular maintenance checks to ensure that the refrigeration system and circulation system are working properly.