Overview
PINGXUAN DC0506-IIX high and low temperature Thermostatic bath is a device that provides accurate and stable temperature environment for scientific research and production. It provides users with a cold and hot controlled, uniform and constant temperature field source through advanced temperature control and refrigeration systems. It is suitable for constant temperature experiments or tests on experimental samples or products, and can also be used as a direct or auxiliary heating/cooling heat source or cold source.
Features
1. Adopt self-developed special software temperature control technology, with High Accuracy PT100 Sensor and imported electronic components, accurate temperature control.
2. The software system supports temperature correction, with resolution up to 0.01 ° C to ensure accurate display temperature.
3. With over-temperature protection and alarm function, the alarm temperature can be set, and the load can be automatically cut off when the temperature is exceeded, which is safe and reliable.
4. Intelligent PID control algorithm can automatically adjust parameters, Temperature Stability, special users can also adjust manually.
5. The software digital locking function can prevent misoperation and ensure that the experimental data is correct.
6. Adopt high-efficiency fully enclosed Compressor for cooling, with fast cooling speed and multiple protections such as overheating and overcurrent.
7. Professional domestic circulation system ensures uniform and stable temperature in the working tank.
8. The power button can turn off all functions, but the instrument can still display the natural process of temperature change in the tank.
Principle
The device monitors the medium temperature in the working tank in real time through the High Accuracy temperature sensor (PT100), and transmits the signal to the special software control system. The control system precisely adjusts the power output of the heating or cooling system (fully enclosed Compressor) through the intelligent PID algorithm according to the difference between the set temperature and the measured temperature. At the same time, the built-in circulating pump drives the medium in the working tank for domestic circulation to ensure the uniformity of temperature distribution, thus achieving fast and stable thermostatic control.
Applications
Research departments of petroleum, chemical industry, electronic instrumentation, physics, chemistry, biological engineering, life science, light industry food property testing, chemical analysis, etc., colleges and universities, enterprise Quality control and production departments
Steps
1. Place the device in a smooth and ventilated place and connect it to the power supply.
2. Add an appropriate amount of heat transfer medium to the working tank.
3. Turn on the power supply and set the required target temperature and over-temperature alarm temperature through the control panel.
4. Start the cooling or heating function, the equipment starts to work, and the domestic circulation system runs synchronously to make the temperature uniform.
5. After the temperature reaches the set value and stabilizes, the sample can be placed for experimentation or testing.
6. After use, press the "Power" button to turn off the function of the device (the device will still display the temperature change in the slot).
7. Disconnect the power supply and clean the working tank.
Notice
• Before use, make sure that the medium liquid level in the working tank is appropriate.
• Do not make unnecessary parameter modifications or misoperation while the device is running.
• Pay attention to the equipment alarm prompt, and check the reason after the overtemperature protection is triggered.
Equipment should be placed in a well-ventilated environment to ensure normal heat dissipation.
Before moving or cleaning equipment, be sure to disconnect the power supply.
• Please manually adjust the PID parameters by professionals as needed.