Overview
PINGXUAN DC3006-IIX high and low temperature Thermostatic bath is a device that provides a precise temperature environment for scientific research and production. It can provide a hot and cold controlled, uniform and constant temperature field source for experimental samples or product testing, and is widely used in Petrochemicals, electronic instrumentation, bioengineering, food physical property testing and chemical analysis and other fields.
Features
1. Adopt self-developed special software temperature control technology, equipped with High Accuracy PT100 Sensor and imported electronic components, accurate temperature control.
2. The software system supports temperature correction, with resolution up to 0.01 ° C to ensure accurate display temperature.
3. With over-temperature protection and alarm function, the alarm temperature can be set, and the load can be automatically cut off when the temperature is exceeded, which is safe and reliable.
4. Intelligent PID control system can automatically adjust parameters, Temperature Stability, special users can also adjust manually.
5. The software digital locking function can prevent misoperation and ensure that the experimental data is correct.
6. Adopt high-efficiency fully enclosed Compressor for refrigeration, with fast cooling speed and multiple protections.
7. Have a professional domestic circulation system to ensure uniform and stable temperature in the tank; at the same time have external circulation function, can establish a second thermostatic field outside the machine or as a cold source.
Principle
The device monitors the temperature in the working tank in real time through the High Accuracy temperature sensor (PT100) and feeds the data back to the intelligent control system. The control system automatically adjusts the output power of the heating or cooling system through the PID algorithm according to the deviation between the set temperature and the measured temperature, so as to quickly and stably control the liquid temperature in the tank at the target value. Domestic circulation pump prompts the liquid flow in the tank to ensure uniform temperature; external circulation pump can output thermostatic liquid for temperature control or chill down of external equipment.
Applications
Petroleum, chemical industry, electronic instrumentation, physics, chemistry, biological engineering, life science, light industry food property testing, chemical analysis, colleges and universities, enterprise Quality control and production departments
Standards
The information does not mention
Steps
1. Turn on the power supply and turn on the device.
2. Set the required temperature value and Miscellaneous parameters through the control panel.
3. Start domestic circulation or external circulation system.
4. The equipment will operate automatically, allowing the temperature in the working tank to reach and stabilize at the set value quickly.
5. After the experiment or test is over, press the "Power" button to turn off the instrument function, but the display screen will still display the natural change process of the temperature in the tank.
Notice
* Please ensure that the equipment is placed in a smooth and well-ventilated environment before use.
Do not allow unrelated personnel to operate the device. You can use the software lock function to prevent misoperation.
* Pay attention to the Temperature range (-30~ 220 ℃) marked on the device, do not use beyond the range.
* When the over-temperature alarm is triggered, please check and troubleshoot in time.
Regularly check the circulating pump and piping to ensure that the circulation system is unobstructed.
* After turning off the power button, the device will still display the temperature. If you need to completely power off, please disconnect the external power supply.