PINGXUAN DC4010-IIX High and Low Temperature Thermostatic bath is a device that provides an accurate and stable temperature environment for scientific research, Quality control and production. It can provide a cold and hot controlled, uniform and constant temperature field source for experimental samples or product testing, and can also be used as a direct or auxiliary heating/cooling heat or cold source.
Features
1. Adopt self-developed dedicated software temperature control technology, equipped with High Accuracy PT100 Sensor and imported electronic components.
2. The software system supports temperature correction, with resolution up to 0.01 ° C to ensure accurate display temperature.
3. With over-temperature protection and alarm function, the alarm temperature can be set, and the load can be automatically cut off when the temperature is exceeded.
4. Intelligent PID control system can automatically adjust parameters, Stability, special users can also adjust manually.
5. The software digital locking function prevents misoperation and ensures that the experimental data is correct.
6. Precise temperature control allows for rapid and stable temperature in the working tank.
7. Adopt high-efficiency fully enclosed Compressor for cooling, with fast cooling speed and multiple protections such as overheating and overcurrent.
8. Have a professional domestic circulation system to ensure uniform and stable temperature.
9. The power button can turn off all functions, but the instrument can still display the natural temperature change process in the tank.
Principle
The equipment uses a self-developed intelligent temperature control system, combined with the High Accuracy PT100 Sensor to monitor the temperature, and automatically adjusts the heating or refrigeration power by the PID algorithm to achieve precise control of the temperature in the working tank. At the same time, its efficient fully enclosed Compressor refrigeration system and professional domestic circulation system work together to quickly reach the set temperature and maintain a high degree of uniformity and stability.
Applications
Research departments of petroleum, chemical industry, electronic instrumentation, physics, chemistry, biological engineering, life science, light industry food property testing and chemical analysis, universities, enterprises Quality control and production departments
Steps
1. Turn on the power supply and turn on the device.
2. Set the required target temperature and over-temperature alarm temperature through the control panel.
3. Start operation, the equipment starts heating or cooling, and the domestic circulation system works to make the temperature uniform.
4. Monitor and display the temperature until it quickly stabilizes to the set value.
5. After use, you can press the "Power" button to turn off the function of the device (the temperature in the slot is still displayed at this time).
6. Disconnect the power supply.
Notice
Make sure the equipment is placed in a smooth and well-ventilated environment before use.
• Do not let unrelated personnel operate the device, and use the software lock function to prevent misoperation.
• Pay attention to the set over-temperature alarm value to ensure that the device operates within a safe Temperature range.
• Regularly check the operating status of the equipment to ensure that the refrigeration system and circulation system are working properly.
• After turning off the function, if you need to completely power off, please disconnect the power connection.