Overview
PINGXUAN Px-0200M-II standard verification High and low temperature Thermostatic bath is a High Accuracy temperature control equipment, mainly used for Measurement department verification of first and second standard mercury-filledthermometer, BECKMAN thermometer, industrial platinum thermistor, standard copper-constantan Thermocouple, can also be used as High Accuracy heat source or cold source for production and scientific research.
Features
1. Self-made temperature control technology developed by dedicated software, equipped with High Accuracy PT100 Sensor and imported electronic components.
2. The software system supports temperature error correction, and the correction resolution is up to 0.001 ° C to ensure that the display temperature is accurate.
3. With over-temperature protection and alarm function, the alarm temperature can be set, and the load can be automatically cut off when the temperature is exceeded.
4. Intelligent Software Temperature Stability, PID can automatically adjust the best parameters according to the medium, and also support manual adjustment.
5. The software digitally locks the control system setting value to prevent unrelated personnel from misusing and ensure that the experimental data is correct.
6. Accurate temperature control allows for rapid and stable temperature in the working tank.
7. Equipped with a fully enclosed Compressor refrigeration system, with fast cooling speed and multiple protections for overheating and overcurrent.
8. Press the "Power" button to turn off all functions of the instrument.
Principle
The device is controlled by a dedicated software technology combined with a High Accuracy PT100 temperature sensor to monitor the temperature in the working tank in real time. The control system automatically adjusts the power output of the heating or cooling system through an intelligent PID algorithm according to the deviation between the set temperature and the measured temperature to achieve fast and accurate temperature control and maintain the uniform and stable temperature in the tank.
Applications
Measurement verification, production, scientific research
Steps
1. Turn on the power supply.
2. Set the required temperature and over-temperature alarm temperature through the control panel or software.
3. Start the device, and the system will automatically control the temperature to reach the set value and maintain stability.
4. Put the thermometer, thermal resistance or Thermocouple to be verified into the working tank for verification or experiment.
5. After use, press the "Power" button to turn off all functions.
Notice
Make sure the equipment is properly grounded before use.
Do not let unrelated personnel operate the equipment to avoid mistakenly changing the setting parameters.
* When the over-temperature alarm function is enabled, do not ignore the alarm prompt.
* When the equipment is running, do not block the ventilating port.
When adding or replacing working media, ensure that it meets the equipment requirements.
Regular maintenance inspections to ensure the normal operation of the equipment.