Overview
PINGXUAN Px-20200M-II standard verification High and low temperature Thermostatic bath is a High Accuracy temperature control equipment, mainly used for Measurement department verification of first and second standard mercury-filledthermometer, BECKMAN thermometer, industrial platinum thermistor, standard copper-constantan Thermocouple, can also be used as High Accuracy heat and cold source for production and scientific research.
Features
1. Using dedicated software to develop self-made temperature control technology, equipped with High Accuracy PT100 Sensor and imported electronic components.
2. The software system supports temperature correction with resolution up to 0.001 ° C to ensure accurate display temperature.
3. With over-temperature protection, over-temperature alarm function, can set the alarm temperature and automatically cut off the load.
4. Intelligent Software Temperature Stability, PID can automatically adjust parameters, special users can also adjust manually.
5. The software digitally locks the control system setting value to prevent unrelated personnel from misusing and ensure that the experimental data is correct.
6. Accurate temperature control allows for rapid and stable temperature in the working tank.
7. Equipped with an effective fully enclosed Compressor for refrigeration, with fast cooling speed and multiple protections for overheating and overcurrent.
Principle
The equipment is controlled by a dedicated software system combined with a High Accuracy PT100 temperature sensor to monitor the temperature in the working tank in real time. The intelligent PID algorithm automatically or manually adjusts the output power of the heating or cooling system according to the deviation of the set temperature and the measured temperature to achieve fast and accurate temperature control and stabilize the temperature near the set target value.
Applications
Measurement verification, production, scientific research
Steps
1. Turn on the power supply and set the required temperature and Miscellaneous parameters through the control system.
2. Start the equipment, the system starts to work, and the temperature in the working tank reaches and stabilizes at the set value.
3. During use, the temperature can be monitored and corrected through software.
4. After use, press the "Power" button to turn off all functions of the instrument.
Notice
• Please read the manual carefully before use to ensure correct operation.
• Set the over-temperature alarm temperature to automatically protect the equipment in case of abnormal conditions.
• Avoid unrelated personnel from operating the control system to prevent parameters from being changed by mistake.
• Pay attention to the operating environment of the equipment to ensure good ventilation.
• Regular maintenance inspections to ensure that the refrigeration system and protection functions are working properly.