Overview
HUXI HT-W330G (set) Multifunctional Benchtop Shaker is a brushless DC motor and microcomputer control technology equipment, with a unique knob operation mode, easy to use. By changing different trays, various commonly used flasks, culture dishes, beakers can be mixed and fostered, suitable for biotechnology, microbiology, medical analysis and other fields. The heating series integrates oscillation and heating functions, and can rotate or reciprocate the sample, mix and foster.
Features
1. Large screen LCD display timing time, Rotation speed, temperature value, Operating interface is simple and easy to understand;
2. Locked operation mode to prevent misoperation;
3. DC brushless motor drive, low noise, accurate speed, long service life, maintenance-free;
4. A variety of pallets, cradles, and fixtures are available for a wide range of needs.
5. Accessories are easy to replace, which greatly improves work efficiency.
6. Soft start, accelerate uniformly, effectively avoid sample splashing;
7. Optional RS232 interface, connected to the computer through the RS232 interface, which can realize the setting of the instrument parameters and the real-time display of the working Linear dispersion on the computer side;
8. The heating series has fast heating speed, wide Temperature range, and the maximum temperature can reach more than 100 ℃, which makes up for the blank that Shaker cannot operate at high temperature.
Principle
Multifunctional Benchtop Shaker is driven by DC brushless motor, using microcomputer control technology implementation of accurate oscillation and temperature control. The device supports reciprocating oscillation, pendulum amplitude is 8mm, Oscillation Frequency range is 60~ 400rpm (optional to 500rpm), by changing different trays to adapt to a variety of containers, to achieve sample mixing, mixing and fostering.
Applications
Biotechnology, microbiology, medical analysis, chromadizing or decolorization of colloids, combinatorial chemistry
Standards
Shell Protection Rating: IP21
Steps
1. Select the appropriate pallet as required and install it.
Connect the power supply to ensure that the voltage is within the range of 100 to 240 volts.
3. Set timing time, Rotation speed and temperature parameters through knob;
4. Start the device and observe the LCD display to ensure that the parameters are correct.
5. Optional RS232 interface to connect to the computer for parameter setting and monitoring;
6. After the operation is complete, turn off the power and clean the equipment.
Notice
• Ensure that the Ambient Temperature is within the range of 5~ 40 ℃;
• relative humidity not exceeding 80%;
• The maximum load does not exceed 7.5KG.
• Power supply voltage check before use;
• Avoid long-term operation in humid or high temperature environments;
• Regular maintenance of motors and pallets to ensure performance;
• Pay attention to high temperature when heating to prevent burns.