Overview
NELE DC-5020 (Vertical) Low Temperature Circulating Bath uses a fluorine-free refrigeration system, the main components are directly imported, and the performance is stable and reliable. It provides a hot and cold controlled, uniform and constant temperature liquid environment for petroleum, chemical, electronic instrumentation, physics, chemistry, biological engineering, life science, light industrial food, physical property testing and chemical analysis research departments, universities, enterprises Quality control and production departments, for direct heating or cooling, and as an auxiliary temperature source.
Features
1. Fully enclosed environmentally friendly Compressor refrigeration system, free of CFCs.
2. Temperature range can be selected from -60~ 100 ° C, and the medium can choose pure water, ethylene glycol, alcohol, etc. according to the temperature.
3. Integrated PID automatic temperature control program, built-in High Accuracy PT100 temperature sensor, large screen LCD display, resolution 0.01 ° C.
4. The capacity range can be selected from 6L to 100L, and non-standard customization is supported.
5. Non-heat centrifugal circulating pump domestic circulation system, porous U-shaped reflux design, to achieve a high degree of uniformity in the flow field and temperature field.
6. The shell is anti-aging static spray, and the liner and countertop are made of S304 brushed stainless steel.
7. Humanized R rounded corner design to avoid operator injuries.
8. The drainage system is convenient for fluid exchange.
9. Multiple Safety protection: power-off protection, upper and lower limit temperature alarm, low water level protection and alarm, temperature runaway protection and alarm, heating device anti-dry burning and explosion-proof, Compressor overheating and overcurrent protection.
Principle
The Low Temperature Circulating Bath achieves precise temperature regulation through a fully enclosed environmentally friendly Compressor refrigeration system and an integrated PID automatic temperature control program, combined with the High Accuracy PT100 Sensor. The domestic circulation adopts a heatless centrifugal pump and a porous U-shaped return design to ensure uniform and stable temperature in the tank and provide a thermostatic liquid environment for the external system.
Applications
Petroleum, chemical industry, electronic instrumentation, physics, chemistry, biological engineering, life science, light industry food, physical property testing, chemical analysis, institutions of higher learning, enterprise Quality control, production department
Notice
• Make sure the equipment is well grounded before use to avoid the risk of electric shock.
• Regularly check the water level to prevent low water levels from causing the heating device to dry out.
• Pay attention to the temperature setting during operation to avoid alarming if it exceeds the range.
• Turn off the power during maintenance to ensure safety.
Temperature range according to select the appropriate medium (such as pure water, ethylene glycol, alcohol).
• Avoid use in humid environments to prevent equipment damage.
• In case of abnormal conditions (such as temperature loss), stop using it immediately and contact a professional.