Overview
NELE GDH-1030 ultra-high accuracy Low Temperature Circulating Bath is a thermostatic device with a fully enclosed environmentally friendly Compressor refrigeration system. Temperature range -10~ 100 ℃, Tank volume 30L, Tank opening dimensions 310 * 280mm are optional. The device has a built-in integrated new generation PID automatic temperature control program and High Accuracy PT100 temperature sensor to ensure accurate and stable temperature control.
Features
1. Adopt imported brand French Taikang or Japanese Panasonic fully enclosed environmentally friendly Compressor refrigeration system
2. Built-in integrated new generation PID automatic temperature control program and High Accuracy PT100 temperature sensor AAA level
3. Equipped with large-screen high-brightness LCD screen, Chinese and English display, digital display resolution 0.0001 ℃
4. Adopt a non-heat centrifugal circulation pump to avoid the traditional water pump from affecting the temperature field of the tank due to its own heat production
5. The domestic circulation system is designed with the principle of porous U-shaped reflux circulation to achieve turbulence and high uniformity in the flow field and temperature field in the tank
6. The fuselage shell adopts anti-aging electrostatic spray technology, and the liner and countertop are made of S304 brushed stainless steel
7. Humanized R rounded corner design structure to avoid operators being injured by edges and corners
8. Safety protection devices include power-off protection, upper and lower limit temperature alarm, low water level protection, temperature loss protection and other functions
Principle
The equipment achieves accurate temperature control through a fully enclosed environmental protection Compressor refrigeration system and a PID automatic temperature Control system. The non-heat centrifugal circulation pump and the porous U-shaped return circulation principle are used to form turbulent flow in the medium in the tank to ensure a highly uniform distribution of the temperature field.
Applications
Laboratories, scientific research institutions, industrial production, mass testing
Notice
• Temperature range according to select the appropriate medium (pure water, ethylene glycol alcohol)
• Pay attention to the low water level protection function of the equipment to avoid dry burning
• Regularly check the operating status of the Compressor
• Keep equipment well ventilated