Overview
NELE GDH-5050 Ultra-high accuracy Low Temperature Circulating Bath is a precision temperature control equipment using a fully enclosed environmentally friendly Compressor refrigeration system. Temperature range covers -50~ 100 ° C, Tank volume is 50L, with High Accuracy Temperature Control and Uniform Temperature Field Characteristics.
Features
1. Adopt imported brand French Taikang or Japanese Panasonic fully enclosed environmentally friendly Compressor refrigeration system
2. Built-in new generation PID automatic temperature control program and High Accuracy PT100 temperature sensor AAA level
3. Equipped with large-screen high-brightness LCD screen, Chinese and English display, digital display resolution up to 0.0001 ℃
4. Design using non-heat centrifugal circulation pump and porous U-shaped return circulation principle
5. The fuselage shell adopts anti-aging electrostatic spray technology, and the liner and countertop are made of S304 brushed stainless steel
6. Humanized R rounded corner design structure to avoid operators being injured by edges and corners
7. Safety protection device: power-off protection, upper and lower limit temperature alarm, low water level protection, temperature loss protection, heating device anti-dry burning and explosion-proof, Compressor overheating and overcurrent multiple protection
Principle
The equipment realizes accurate temperature control through fully enclosed Compressor refrigeration system and PID automatic temperature control program, adopts the principle of non-heat centrifugal circulation pump and porous U-shaped return circulation to ensure high uniformity of flow field and temperature field in the tank, and has built-in High Accuracy PT100 temperature sensor to monitor temperature changes in real time.
Applications
Laboratory temperature control experiment, scientific research institutions temperature test, industrial production process temperature control, mass detection temperature environment simulation
Standards
No relevant information yet
Steps
1. Select the appropriate medium (pure water, ethylene glycol alcohol) according to the Temperature range
2. Check the water level of the equipment to ensure that it is within a safe range
3. Set the required temperature parameters and alarm upper and lower limits
4. Start the device and observe the temperature stability
5. Turn off the device according to the specified procedures after use
Notice
• The water level must be checked before use to avoid low water level operation
• Select the appropriate medium strictly according to the Temperature range
• Pay attention to the temperature alarm prompt when the equipment is running
• Avoid dry burning of the heating device at low water level
• Regularly check the operating status of the Compressor
• Pay attention to the edges and corners of the equipment during operation to avoid accidental injuries