Overview
BANTE S30-Cn is a high-performance wireless Bluetooth ion meter dedicated to the measurement of cyanide Ion concentration. The device adopts Bluetooth 4.0 technology, is compatible with Android smartphones or tablets, supports 5 point button calibrate, can be equipped with various ion selective Electrode, Measurement accuracy is +/- 1% F.S, range is 0.000005~ 0.01mol/L.
Features
1. High-performance Bluetooth ion meter, providing 5 Ion concentration Measuring methods.
2. Optional Ion concentration units, including ppm, mg/L, mol/L and mmol/L.
3.2 to 5 point calibrated, including 8 concentration calibrated points optional.
4. Electrode management, can store 3 Electrode slope and call calibrate data.
5. Manual temperature compensation to eliminate measurement errors caused by temperature deviation.
6. Automatic end point locking, maintaining stable readings for easy browsing and recording.
7. Timed interval readings, sending measurement data to memory or printer.
8. Limit alarm, automatic warning Measured value exceeds the set range.
9. Calibrate expiration reminder prompts the user to calibrate the meter regularly.
10. calibrate report, showing date, time, calibrate point and slope.
11. Password protection against unauthorized calibrating with settings.
12. Multi-Parameter measurement, can connect Bluetooth conductivity and dissolved oxygen meter to the device and display Measured value.
13. Reset function, automatically restore the meter to the factory default settings.
Principle
Measurement range of +/- 2000.0mV with accuracy of +/- 0.2mV. The ion meter supports manual temperature compensation (0~ 100 ° C/32~ 212 ° F) to eliminate temperature effects and offers multiple Measurement modes (e.g. Continuous or Automatic End Lock) and Stability conditions (Fast, Standard, High Accuracy) to ensure accurate reading of Cyanide Ion concentrations.
Steps
1. Turn on the meter and ensure that Bluetooth is connected to the Android device.
2. Select Cyanide Ion Measurement mode.
3. Perform 2 to 5 point calibrating with optional calibrating fluid (e.g. 0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000).
4. Measurement unit (ppm, mg/L, mol/L or mmol/L).
5. Enable manual temperature compensation as required.
6. Start the measurement, you can choose automatic end lock or continuous mode.
7. View and record data, support sending to memory or printer.
Notice
• Regularly calibrate the meter and set a calibrate reminder (1 to 99 days) to ensure accuracy.
• When using manual temperature compensation, ensure that the temperature setting is accurate.
• Measurement range (0.000005~ 0.01mol/L) to avoid exceeding the range.
• Keep the Bluetooth connection within 10 meters.
• Use password protection to prevent unauthorized operation.
• Use 1.2V lithium battery or 2 AAA batteries when replacing the battery.