Overview
The TENLIN SC-15A High Accuracy Thermostatic bath is a precision temperature control device with a Tank volume of 15L with a Temperature range from Room Temperature to 200 ° C. It uses an advanced Refrigeration and Control system designed to provide a stable, uniform thermostatic environment for laboratory and industrial applications.
Features
1. Low Noise Air-Cooled Fully Enclosed Compressor Unit, Fast Cooling Speed and Strong Temperature Stability.
2. Equipped with self-developed dedicated software and High Accuracy Pt100 Sensor, combined with all imported electronic components, to achieve precise temperature control.
3, with internal and external circulation system: external circulation can establish a second thermostatic field or as an external cold source; domestic circulation to ensure that the liquid temperature in the tank is highly uniform and stable.
4. It has a complete protection system, including cooling overheating/overcurrent protection, over-temperature sound and light alarm, and over-temperature automatic load cutting.
5. Adopt high-end large LCD control instrument, easy to operate, and support P.I.D automatic search or manual adjustment.
6. Temperature correction accuracy up to 0.01 ° C, Temperature Fluctuation up to +/- 0.01 ° C (depending on the model).
Principle
The Thermostatic bath works in tandem with the refrigeration system (fully enclosed Compressor unit) and the heating system to achieve rapid arrival and precise maintenance of the target temperature. The Control system monitors the liquid temperature in the tank in real time through the High Accuracy Pt100 Sensor and automatically adjusts the cooling or Heating power using the P.I.D. algorithm. The circulation system (domestic circulation or external circulation) ensures that the temperature of the working medium is evenly distributed, thus providing a highly stable thermostatic field for the sample being tested or external equipment.
Applications
Laboratory thermostatic experiments, industrial process calibrating, material performance testing, biochemical fostering, precision instruments
Standards
The data did not mention specific compliance criteria
Steps
1. Place the device in a stable and well-ventilated position and connect the power supply.
2. Inject an appropriate amount of suitable medium (e.g. silicone oil, water, etc.) into the Tank.
3. Turn on the power switch and set the required target temperature, alarm temperature upper and lower limits, and other parameters through the large LCD control panel.
4. Choose domestic circulation or external circulation mode as needed.
5, start operation, the equipment will automatically cool or heat to the set temperature and remain thermostatic.
6. After use, turn off the power supply of the equipment and do a good job of cleaning and maintenance.
Notice
Equipment should be placed in a well-ventilated, non-corrosive gas environment.
• Ensure that the medium liquid level in the Tank is appropriate to avoid dry burning or overflow.
• When setting the over-temperature alarm temperature, the upper and lower limits should be set reasonably to ensure safety.
• When making external circulation connections, ensure that the pipeline connection is firm to prevent leakage.
• Regularly clean the surface and tank of the equipment to keep it clean.
• If not used for a long time, the medium in the tank should be emptied and properly stored.