Overview
HUXI LED overhead Stirrer (model: RWD50E/20030037) is a laboratory equipment suitable for mixing medium and high viscosity liquids. The product is capable of high load, long continuous operation, outstanding performance, and automatic protection for overload and abnormal conditions to ensure laboratory safety.
Features
1. LED display, Rotation speed can be coarse and fine-tuned, and Rotation speed can also be adjusted during the operation of the instrument.
2. The motor starts softly and runs smoothly, effectively preventing sample overflow.
3. The closed enclosure effectively prevents liquid from splashing into the instrument, ensuring the safe operation of the experiment.
4. Configure the Impeller penetration hole, the user does not need to adjust the installation height of the instrument, just adjust the installation position of the stirring rod.
5. Provide a variety of Impellers and brackets to choose from.
Principle
The Impeller swirl/spin motor is used to mix medium and high viscosity liquids. With stepless speed regulation, the user can precisely adjust the Rotation speed as needed. Its enclosed housing and soft start design ensure smooth and safe operation.
Applications
Chemical synthesis, physical and chemical analysis, petrochemical, cosmetics, health products, food, biotechnology
Standards
Protection Rating: IP42
Steps
1. According to the characteristics of the liquid to be stirred, select the appropriate Impeller and install it.
2. Adjust the installation position of the stirring rod through the Impeller penetration hole.
3. Turn on the power supply.
4. Rotation speed is coarse and fine-tuned through the LED screen and knob, and the instrument can also be adjusted during operation.
5. Start the equipment to stir.
Notice
● Power supply voltage in 100-240VAC range.
● Ambient Temperature of the equipment is 5~ 40 ℃, relative humidity ≤ 80%.
Do not exceed the maximum Processing capacity (20L) and maximum viscosity (10000mPa · s) of the equipment.
● Note that the diameter range of the drill chuck clamping is 0.5 to 10mm.
The equipment is equipped with overload and motor protection functions, and will issue an audible prompt and automatically stop. In this case, the load should be checked and troubleshooted.
● Use non-slip pads to ensure stable placement of the equipment.
● Impeller need to use matching wrench chuck fixed firmly.