Overview
HUXI LED overhead Stirrer (model: RWD100E/20030045) is a laboratory equipment suitable for mixing medium and high viscosity liquids. The product is capable of high load, long continuous operation, outstanding performance, and automatic protection for overload and abnormal conditions to ensure laboratory safety.
Features
1. LED screen display, Rotation speed can be coarse and fine-tuned, and Rotation speed can also be adjusted during the operation of the instrument.
2. Motor soft start, smooth operation, effectively prevent sample overflow.
3. The closed enclosure effectively prevents liquid from splashing into the instrument, ensuring the safe operation of the experiment.
4. Configure the Impeller penetration hole, the user does not need to adjust the installation height of the instrument, just adjust the installation position of the stirring rod.
5. Provide a variety of Impellers and brackets to choose from.
Principle
The Stirrer uses a motor to drive the Impeller swirl/spin to mix medium to high viscosity liquids. Its motor has a soft start function to ensure smooth operation. The instrument is safe during the experimental process through a closed case and an automatic protection mechanism.
Applications
Chemical synthesis, physical and chemical analysis, petrochemical, cosmetics, health products, food, biotechnology
Standards
Protection Rating: IP42
Steps
1. Select the appropriate Impeller according to the sample viscosity and container.
2. Install the stirring rod through the Impeller penetration hole and adjust it to the appropriate position.
3. Use a wrench chuck to secure the Impeller.
4. Turn on the power supply and adjust the Rotation speed coarsely or finely through the LED screen and knob.
5. Start the equipment, and adjust the Rotation speed as needed during operation.
Notice
• Please ensure that the Ambient Temperature is used under the conditions of 5~ 40 ℃ and relative humidity ≤ 80%.
• Power supply voltage 100-240VAC, 50/60HZ.
• Impeller clamping diameter range is 0.5~ 10mm, please do not use beyond the range.
• The device has an overload protection function, which will issue a sound prompt and automatically stop when triggered. Please check the cause before using it.
• Pay attention to the non-slip pad at the bottom of the instrument to ensure smooth placement.
• When the instrument is running, do not put your hands or foreign objects near the swirl/spin parts.