Overview
SANSHEN YM100L Touchscreen Controlled Vertical Pressure Steam Autoclave is a vertical sterilizing device made of stainless steel and equipped with a radial self-expanding sealing ring. It has an upper touch LCD screen, supports microcomputer custom program setting, the design pressure is 0.25Mpa, the design temperature is 138 ° C, the Operating temperature can be set arbitrarily in the range of 105-135 ° C, and the volume is 100L. It is a simple Pressure Vessel that does not require a license according to regulations.
Features
1. Stainless steel material, radial self-expanding sealing ring.
2. Dial rod elastic cover structure, automatic bounce when opening the cover, and the upper cover has a high temperature and anti-scalding protective structure.
3. The upper touch LCD screen is convenient for operation and observation.
4. Microcomputer custom program settings to meet different sterilizing needs.
5. The large screen intuitively displays the sterilizing process and displays the running status of the equipment.
6. Pressure safety interlock device, the upper cover cannot be opened under pressure.
7. Automatic water replenishment function for water shortage. After the water replenishment is completed, the heating will be automatically resumed.
8. With sound and light alarm system, water shortage, sterilizing end beep prompt.
9. Over-temperature automatic control and safety valve overpressure self-leakage double protection.
10. Ultra-quiet design, with its own exhaust gas treatment system, steam domestic circulation.
11. Cold air automatic discharge function during sterilization.
12. There is automatic exhaust or manual exhaust at the end of sterilizing.
13. Real-time printing function, record sterilizing process data; and a USB port, you can sterilize the record copy U disk save.
14. It has the function of viewing the number of times the machine is used.
15. Built-in TT/PT temperature, stress testing verification interface.
16. Universal casters with brakes, easy to shift and easy to orient.
17. Immersion electric heating tube, high thermal efficiency.
Principle
The Autoclave generates high-temperature steam by heating water with an immersion electric heating tube, and uses a microcomputer Control system to precisely control the temperature, pressure and time of the sterilizing process. During the sterilizing process, the cold air is automatically discharged to ensure that the sterilizing chamber is filled with saturated steam to achieve the sterilizing effect. After sterilizing, the steam can be discharged automatically or manually.
Applications
Medical, laboratory, scientific research, pharmaceutical, biotechnology
Standards
Standard for Simple Pressure Vessels
Steps
1. Connect the power supply (AC220V/50HZ) and the water supply (pressure ≥ 0.2Mpa).
2. Set the required sterilizing temperature (105-135 ° C) and sterilizing time (≤ 999min) through Touchscreen.
3. Place the items to be sterilized and close the Autoclave cover.
4. Start the sterilizing program, and the equipment will automatically complete the process of replenishing water, heating, exhausting and sterilizing.
5. After sterilizing, you can choose automatic exhaust or manual exhaust.
6. After the pressure drops to a safe range, open the upper cover and remove the item.
7. Export sterilizing process data via real-time printing or USB port.
Notice
• The equipment is a simple Pressure Vessel and does not require a license, but it still needs to be used in accordance with the operating procedures.
• Ensure that the replenishment pressure is not less than 0.2Mpa to ensure the normal function of automatic replenishment.
• When the Autoclave is under pressure, the safety interlock device will lock the upper cover and forcibly open it is strictly prohibited.
• Please pay attention to high temperature parts during operation to avoid burns.
• Regularly check whether the safety valve and other protective devices are normal.
• When moving your device, use a universal caster with brakes and pay attention to safety.