Overview
SANSHEN YM100LII vertical pressure steam Autoclave is a microcomputer program-controlled sterilizing equipment with a volume of 100L. It is made of S30408 stainless steel. Operating temperature can be set arbitrarily in the range of 105-136 ° C, and the temperature control accuracy can reach +/- 0.5 ° C. It is suitable for high temperature and high voltage steam sterilization of items in laboratories, medical and other fields.
Features
1. Using S30408 stainless steel material and radial self-expanding silicone sealing ring.
2. Dial rod elastic cover structure, automatic bounce when opening the cover, and the upper cover has high temperature and anti-scalding protection.
3. The control display panel is placed on the flip cover for easy operation and observation.
4. Microcomputer program control, providing a variety of custom program options.
5. LED digital tubes and indicator lights dynamically display the operating status of the device.
6. The pressure gauge simultaneously displays the pressure and temperature in the barrel.
7. The computer board has PID function, and the heating parameters are corrected by self-tuning, and the temperature control accuracy is within +/- 0.5 ° C.
8. Equipped with a pressure safety interlock device, the upper cover cannot be opened under pressure.
9. With low water level control function, automatic water replenishment when the water level is insufficient, and automatic recovery of heating after water replenishment.
10. Automatically stop heating and sound and light alarm when the temperature exceeds the set value of 2 ° C.
11. Built-in TT/PT temperature and stress testing verification interface.
12. Equipped with universal casters with brakes, it is easy to move.
13. There is a buzzer prompt at the end of sterilizing.
14. It is a simple Pressure Vessel and can be used without a license according to regulations.
Principle
The Autoclave generates high temperature and high voltage saturated steam by electric heating. The microcomputer program controls the heating process, so that the sterilizing chamber reaches and maintains a set temperature (such as 121 ° C or 134 ° C) and pressure (Operating Pressure 0.23Mpa) for a certain period of time. Under this condition, the saturated steam can effectively kill all microorganisms including bacterial spores, and realize the sterilization of items. After sterilizing, the system automatically stops heating.
Applications
Laboratories, medical and health institutions, biopharmaceuticals, food inspection, scientific research institutes
Standards
Standard for Simple Pressure Vessels
Steps
1. Check the equipment status to ensure that the water level is normal.
2. Put the sterilized items into the sterilizing basket and put them in the sterilizing room.
3. Close the upper cover and ensure sealing.
4. Set the required sterilizing temperature, time and other parameters through the control panel.
5. Start the sterilizing program, and the equipment will automatically complete the heating, sterilizing and heat preservation process.
6. After sterilizing, the buzzer prompts, after the pressure drops to a safe range, open the upper cover and take out the item.
Notice
Please read the instructions carefully before use.
• Ensure that the replenishment pressure is between 0.2Mpa and 0.3Mpa.
• When Autoclave is under pressure, it is strictly forbidden to forcibly open the upper cover.
• The items placed should not be too full, and a gap should be left to ensure steam circulation.
• When taking out sterilized items, pay attention to high temperature and anti-scalding.
• Regularly check whether safety devices such as safety valves are normal.
Equipment should be placed in a ventilated, drying and stable place.
Before performing maintenance, be sure to cut off the power supply.