Overview
YIHENG THZ-35 thermostatic foster Shaker combines more than ten years of design and manufacturing experience, introduces and digests foreign technologies, and is guided by user requests. It is widely used in cell fostering, fermentation, hybridization, biochemistry and cell tissue research that require high temperature and Oscillation Frequency. It can carry out movement and static fostering of microbial cells and various strains.
Features
1. Humanized design: set incubator, Shaker in one, small footprint; Large LCD screen, real-time display temperature, Rotation speed and Working time, Menu Operating interface is easy to understand; Fluorine-free design and environmental protection; Large observation window, the inner tank of the box is made of 304 stainless steel, Shaker Table surface is treated with high-grade aluminum surface bacteriostatic coating, which is easy to clean.
2. Quality assurance: Microcomputer control temperature and Oscillation Frequency, with timing function; built-in power-off protection function, automatic recovery operation after normal power supply; key components such as Compressor and circulating fan use brand products, environmentally friendly without fluorine refrigerant.
3. Continuous operation guarantee: low heat dissipation brushless DC motor, large starting torque, wide speed regulation, maintenance-free, can run continuously for a long time.
4. Safety function: equipped with door switch, automatic stop breeze circulation, heating and Shaker when the box door is opened; unique control Rotation speed circuit to ensure smooth start and prevent liquid splashing; independent temperature limit alarm system, automatic cut off heating over limit; automatic control of circulating fan speed to avoid sample volatilization; optional RS-485 interface and USB data transfer interface, connecting computer to monitor temperature, Rotation speed, time and alarm.
Principle
The thermostatic foster Shaker combines the thermostatic environment with oscillatory motion by precisely controlling the temperature and Oscillation Frequency with a microcomputer controller. The built-in brushless DC motor drives the Shaker Table, cooperates with a Compressor and a circulating fan to maintain temperature stability, ensuring that the sample is fostered under set conditions.
Applications
Cell fostering, fermentation, hybridization, biochemistry, cell tissue research, microbial cell and strain fostering
Standards
Fluorine-free design environmental standards
Steps
1. Connect the power supply and turn on the device.
2. Set the Temperature, Oscillation Frequency and Working time through the large LCD menu.
3. Place the sample on the Shaker Table and fix it with a fixture.
4. Start the device and monitor the running status.
5. After running, turn off the equipment and clean it.
Notice
• Automatically stop running when the box door is opened to avoid temperature overshoot.
• Ensure that the oscillation bearing weight does not exceed 10Kg (including the fixture).
• Power supply voltage is 220V 50Hz before use.
• Regularly clean the cabinet liner and Shaker Table to maintain hygiene.
• Automatically cut off heating when the temperature exceeds the limit, pay attention to the alarm prompt.
• Optional interface to connect to computer monitoring data, which needs to be configured correctly.