Overview
YOKE GT124C is an internal calibration Touchscreen Balance, which belongs to the I-level accuracy class. It uses ARM32-bit microprocessor chip, combined with digital filtering and excellent data processing technology to ensure fast Response and high Measuring accuracy. The instrument uses excellent electromagnetic Sensor technology and High Accuracy numerical control processing accessories to ensure the accuracy and Stability of the Sensor, and is equipped with sealing Draft Shield, which is beautiful in appearance and easy to use.
Features
1. Built-in ARM32-bit microprocessor chip, fast running speed and high Measuring accuracy.
2. Adopt digital filtering and excellent data processing technology to intelligently improve Response Speed and Precision of measurement data.
3. Using excellent Electromagnetic Sensor technology, Sensor accessories are processed by High Accuracy numerical control machining center to ensure accuracy and Stability.
4. Adopt sealing Draft Shield, beautiful appearance and easy to use.
5. Contains counting, weighing, density, animal weighing and other modes to meet different needs, simple operation and high Precision.
6. The circuit board adopts SMT chip process, which improves the Reliability of the product.
7. BTN black film LCD screen with white backlight, clear display.
Principle
The Balance is weighed based on Electromagnetic Sensor technology. The built-in ARM32-bit microprocessor chip processes the Sensor signal and intelligently processes the measurement data through digital filtering and excellent data processing technology to output results quickly and accurately. Its internal Adjustment automates the calibration process through built-in test weights and programs.
Applications
Counting and weighing, checking and weighing, density measurement, animal weighing
Standards
Class I accuracy class
Steps
Place the Balance on a smooth countertop and connect it to a 220V/50Hz power supply.
2. Turn it on to warm up and wait for stability.
3. Internal or external calibrating as required.
4. Select the desired weighing mode (e.g. count, check, density, animal weighing) through the Touchscreen.
5. Place the item to be weighed on a φ 90mm pan for weighing.
6. Read and record the stable data displayed on the LCD screen.
Notice
• Please place the Balance on a smooth, vibration-free countertop for use.
• Ensure that the Operating Environment meets product requirements and avoids strong airflow interference.
Before calibrating, make sure the Balance is fully warmed and stable.
• Do not exceed 120g Capacity.
• When cleaning, use a soft dry cloth to avoid liquids entering the inside of the instrument.
• Power supply voltage is 220V/50Hz, do not use a mismatched power supply.