Overview
YOKE P611 acidity meter is a P6 series Portable pH measuring instrument with large screen LCD display and white backlight. It has a built-in microprocessor, integrates various intelligent functions such as Automatic calibration, temperature compensation, Data storage, and meets IP57 waterproof rating. With a suitcase, it is suitable for field and field use.
Features
1. Large LCD display, white backlight controllable.
2. Built-in microprocessor chip, with automatic calibration, automatic temperature compensation, data storage, function settings, self-diagnostic information, automatic shutdown and low voltage display and other intelligent functions.
3. Automatically identify 13 kinds of pH standard buffer solutions, and choose from European, American, NIST, and Chinese series.
4. High purity water pH Measurement mode and ammonia pure water pH Measurement mode can be set.
5. Adopt bracket design and non-slip glove design for easy use and operation.
6. Configure calibrated solution and suitcase, suitable for field and field use.
The instrument meets the IP57 waterproof rating.
Principle
The instrument uses pH Electrode to measure the activity of hydrogen ions in the solution and convert it to ph value. The built-in microprocessor chip is responsible for processing Electrode signals and correcting the measurement results with automatic temperature compensation to ensure measurement accuracy at different temperatures. The data can be stored and displayed on the LCD screen.
Applications
Field measurement, field testing, laboratory analysis, water quality monitoring, industrial process control
Standards
ISO9001:2000、CE、CMC
Steps
1. Install the battery and turn it on.
2. Select the standard buffer solution series (Europe, America, NIST or China series) according to the measurement needs and perform Automatic calibration.
3. Measurement mode of high purity water or ammonia pure water can be set for measurement of special water sample.
4. Immerse Electrode in the solution to be tested and wait for the reading to stabilize.
5. Read and record the pH value, temperature or mV value displayed on the LCD screen.
6. The measurement data can be automatically stored.
7. After use, clean the Electrode according to the instructions and shut down.
Notice
• Please ensure that the instrument is used at an Ambient Temperature of 5~ 35 ° C and ≤ 85% Environmental humidity.
• When calibrating and using, use the specified standard buffer solution to ensure measurement accuracy.
• Pay attention to the battery level, a prompt will be displayed when the voltage is low.
• Although the instrument has an IP57 waterproof rating, it is still necessary to avoid long-term immersion or extreme environments.
• Electrode maintenance is essential, please follow the instructions to clean and store.
• The stored data is number, Measured value, Measurement unit and temperature value, a total of 150 groups.