Ford-2 Ford viscosity cup is a Benchtop Ford viscosity cup designed by YONGLIDA brand according to the American ASTM D1200 standard, which is widely used in Europe, America and other countries. This model is specially designed for laboratory viscosity measurement Coatings, paints, inks and other viscosity, with specific aperture and Measurement range.
Features
1. Made of aluminum alloy cup body and stainless steel aperture to ensure durability and accuracy
2. Five models are available, and the outflow aperture of the Ford-2 model is 2.53mm.
3. Measurement range of 25~ 120cSt, suitable for medium viscosity fluids
4. The standard volume is 100ml, which meets the requirements of routine laboratory testing
5. Reliability of measurement results based on ASTM D1200 standard
Principle
Ford viscosity cup calculates viscosity by measuring the time it takes for a specific volume of fluid (100 ml) to flow out of a standard pore size (2.53mm). Flow time is proportional to fluid viscosity and follows ASTM D1200 standard test method.
Applications
Coatings, paints, inks
Standards
ASTM D1200
Steps
1. Place the Ford viscosity cup vertically over the fluid to be tested
2. Pour 100ml of test fluid until the cup is full
3. Start the Timer at the same time and let the fluid flow naturally from the 2.53mm aperture
4. Record the time required for complete fluid outflow
5. Calculate the viscosity value based on the outflow time
Notice
• Make sure the cup body and aperture are clean and drying before use
• Keep Ford viscosity cup vertically placed during testing
• Ambient Temperature should be kept stable to avoid affecting measurement results
• Only suitable for fluids in the specified viscosity range (25~ 120cSt)
• Regularly check the aperture for blockage or abrasion
Viscosity Range 25~120 cSt
flow hole diameter 2.53mm
capacity 100 ml
Cup body material aluminum alloy
Nozzle material stainless steel
Applicable standard ASTM D1200
certificate without
[Q] Is the price listed on the website the final cost for overseas purchases?
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[Q] What is the specific process for purchasing the product?
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[Q] What should I do if the instrument I purchase needs repair?
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.