Please contact customer service
Volume 408L, Temperature range -70~ 150 ℃, humidity control 20% -98% RH. Adopt SUS304 stainless steel studio, vacuum tempered Glass observation window, turbine centrifugal forced air supply and imported Compressor to ensure uniform and stable temperature and humidity. Equipped with a large screen touch controller, support program editing and remote control.
Category : Temperature Humidity Chamber
Brand : TaiSiTe
SKU : NB040955
Service :The delivery and service By NBCLab.Com
TAISITE WSS-408P constant temperature and humidity test box is a device for heat resistance, cold resistance, dry resistance and moisture resistance test. Its internal volume is 408L, the use Temperature range can reach -70 ℃ to 150 ℃, the humidity control range is 20% to 98% RH, with precise temperature and humidity control ability.
1. SUS304 stainless steel studio, equipped with flat reaction handle. 2. Vacuum tempered Glass observation window, with defrosting function, easy to observe the experimental situation at any time. 3. Built-in independent steam humidification system, uniform humidity. 4. Using turbine centrifugal forced air supply, the temperature is more uniform. 5. Using the original imported Compressor, environmentally friendly R404 refrigerant, with spray liquid chill down and spray liquid defrost function, stable and effective refrigeration. 6. Built-in pulley, easy to move. 7. Imported large-screen LCD touch controller, easy program editing, support Chinese and English, real-time display of temperature operation Linear dispersion, equipped with USB3.0 or RS485 interface, support remote control of the experimental process. 8. Easy to operate, support fixed-value operation, regular operation, and automatic stop. 9. Dedicated function keys to achieve temperature setting. 10. Auxiliary menu can achieve over-rise alarm, deviation correction, menu lock. 11. Equipped with power-off protection, parameter memory, and call recovery functions. 12. Compressor overload protection. 13. Support appointment start stop. 14. With loop self-diagnosis, Sensor fault alarm, upper and lower limit over-temperature alarm function. 15. Intelligent defrosting, monitoring timing, overlift preventer. 16. High safety, including over-lift alarm, menu lock, over-lift prevention, overload protection, and power failure protection.
The equipment is heated by a nickel-chromium alloy electric heating wire heater and humidified by a stainless steel sheathed heater through a multi-stage program temperature control system. The refrigeration system adopts a French Taikang fully enclosed Compressor or a German BITZER semi-enclosed Compressor, and cooperates with environmentally friendly R404A and R23 refrigerants to achieve air-cooled single-stage or cascade refrigeration. Through a centrifugal fan and a broadband forced air circulation (up and down), the temperature and humidity in the box are ensured to be uniform. The humidity control adopts an imported wet and dry ball Sensor, and steam humidification is carried out by lifting the water supply through a water pump.
Electronics and electrical engineering, automotive parts, aerospace, materials research, chemical industry, Biomedical, scientific research institutes
Not mentioned
Connect the power supply (AC220V +/- 10% or AC380V/415V +/- 10%). 2. Parameter setting is carried out through the large-screen LCD touch controller, and fixed-value operation or program operation mode can be selected. 3. For remote control, it can be connected via USB 3.0 or RS485 interface. 4. Put the sample to be tested on the shelf of the inner box. 5. Start the equipment and start the experiment. 6. During the experiment, real-time observations can be made through the observation window. 7. After the experiment, the device can be automatically stopped according to the setting.
• Power supply voltage meets equipment requirements (AC220V +/- 10% or AC380V/415V +/- 10%). The equipment is equipped with multiple safety protections (e.g. over-temperature, anti-dry burning, water shortage, overload, etc.), but it still needs to be inspected regularly. • Check the water supply system before use to ensure that the humidification function is normal. • Use the built-in pulley when moving your device and keep it steady. • Do not block incoming or outgoing trends when the device is running. • After conducting low temperature or high humidity experiments, please pay attention to defrosting and dew condensation in the box. • Non-professionals should not disassemble or modify the internal circuits and refrigeration systems of the equipment. • Equipment lead hole (inner diameter 52mm) can be used to test the wire passing through, please pay attention to sealing.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.