Please contact customer service
800L large-capacity studio, using SUS304 stainless steel liner. Temperature control accuracy up to +/- 0.2 ℃, humidity deviation +/- 1.5% RH. Equipped with imported Compressor and turbine centrifugal forced air supply system to ensure uniform and stable insulation field.
Category : Temperature Humidity Chamber
Brand : TaiSiTe
SKU : NB040960
Service :The delivery and service By NBCLab.Com
The TAISITE WSS-800P constant temperature and humidity test chamber is a device for testing heat, cold, dry and humidity environments. With a temperature control range of -70 ° C to 150 ° C and a humidity control range of 20% to 98% RH and an inner product of 800L, it is suitable for a wide range of Reliability testing needs.
1. The studio is made of SUS304 stainless steel, equipped with flat reaction handle and vacuum tempered Glass observation window for easy observation. 2. Built-in independent steam humidification system, with turbine centrifugal forced air supply to ensure uniform temperature and humidity. 3. Using the original imported Compressor and environmental protection R404 refrigerant, with liquid spray chilling down and liquid spray defrosting function, refrigeration is stable and effective. 4. Built-in pulley, easy to move. Equipped with imported large-screen LCD touch controller, supports Chinese and English, program editing, real-time Linear dispersion display, and has USB3.0 or RS485 interface for remote control. 6. Easy to operate, support fixed-value operation, regular operation, automatic stop, appointment start stop and other functions. 7. With multiple safety protection functions, including over-lift alarm, menu lock, over-lift prevention, overload protection, power-off protection, parameter memory, call recovery, loop self-diagnosis and Sensor fault alarm, etc.
The equipment is heated by a nickel-chromium alloy electric heating wire heater and humidified by a stainless steel sheathed heater. The refrigeration system uses a French Taikang or German BITZER Compressor to carry out single-stage or cascade refrigeration with environmentally friendly refrigerants R404A and R23, and dissipates heat through air cooling. The equipment realizes broadband forced air circulation (up, out and down) through a centrifugal fan to ensure uniform temperature and humidity in the box. The control system collects data through a PT100 temperature sensor and an imported wet and dry bulb humidity Sensor, and the touch screen controller performs multi-stage program temperature control to achieve accurate temperature and humidity environment simulation.
Reliability test of electronic products, weathering resistance test of materials, environmental adaptability assessment of parts, mass inspection of products
Not mentioned
Connect the power supply (AC220V +/- 10% or AC380V/415V +/- 10%). 2. Set the required temperature and humidity program or select the fixed value/timing operation mode through the large-screen LCD touch controller. 3. Start the equipment, the system will automatically control the functions of heating, humidification, cooling and defrosting to achieve the set environment. 4. The experimental status can be observed in real time through the observation window, and the controller will display the real-time temperature operation Linear dispersion. 5. After the experiment is completed, the device can be automatically stopped according to the setting, or manually stopped.
• Power supply voltage meets equipment requirements (AC220V +/- 10% or AC380V/415V +/- 10%). The equipment is equipped with multiple safety protections, but it is still necessary to pay attention to prevent overheating, dry burning, water shortage, etc. when using it, and regularly check fuses and other protective devices. • When moving the device, the built-in pulley can be used, but it is necessary to ensure that the ground is flat. • The humidification system needs to ensure the normal water supply and pay attention to liquid level protection. • Non-professionals should not disassemble or modify the internal circuits and refrigeration systems of the equipment. • When the equipment is running, do not open and close the door frequently, so as not to affect the uniformity of temperature and humidity and the test results. • When not in use for a long time, it is recommended to disconnect the power supply and do a good job of cleaning and maintenance.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.