Using Three-Box independent damper switching technology, rapid temperature Impact can be achieved without moving the sample, the temperature recovery time is only 5 minutes, and the test range covers -60 ° C to 150 ° C, meeting the demanding environment Reliability verification requirements.
Category : Thermal Shock Test Chamber
Brand : HUDA
SKU : NB041394
Service :The delivery and service By NBCLab.Com
HUDA QD-E703-50KS40 Thermal Shock Test Chamber is a device used to test the performance of products in a fast temperature alternating environment. It is suitable for aerospace products, information electronic Instruments, materials, electrical, vehicles, metals, electronic products, various electronic components, wires, cables, etc., for inspection of their various performance indicators and mass management.
1. Adopt the independent structure of the pre-cooling zone, the pre-heating zone, and the test zone Three-Box, and achieve rapid temperature Impact through damper switching without moving the test sample. 2. Equipped with imported LED digital display microcomputer integrated controller, using human-machine interaction touch input, easy to operate. 3. The program capacity is large, the maximum can be set to 120 groups of programs, and can realize the use of inter-program connection, each command can be repeability up to 9999 times. 4. Equipped with a variety of control functions, including high temperature/low temperature/normal temperature start, preheating and precooling, program jump hold, PID parameter setting, power-off memory continuation, appointment start-stop, fault recording, etc. 5. Adopt 7-inch LCD color display, which can display temperature setting value, actual value, program Linear dispersion, running status and other information in real time, and support traditional Chinese/English language switching. 6. The refrigeration system adopts European imported low temperature Compressor and binary cascade refrigeration, and the heating system adopts imported nickel-chromium alloy electric heater. 7. Safety protection device, including Compressor overheating/overcurrent/overpressure protection, adjustable overtemperature protection, phase leakage protection, Ambient Temperature protection. 8. Provide RS-485 Communication interface and U disk storage function for data recording, export and remote monitoring.
The equipment is divided into three independent boxes: the pre-cooling area, the pre-heating area and the test area. During the temperature impact, the airflow path is switched by controlling the opening and closing of the damper: when the impact is at room temperature, the damper of the high and low temperature groove is closed, and the Ambient Temperature is introduced into the test area through the blower; when the impact is at low temperature, the damper of the high temperature and normal temperature grooves is closed, so that the low temperature groove is connected with the test area, and the pre-stored heat is introduced in an instant. Rapid alternation of temperature is achieved in this way.
Aerospace Products, Information Electronic Instruments, Materials, Electricians, Vehicles, Metals, Electronic Products, Various Electronic Components, Wires, Cables
GJB150.5A-2009 temperature impact test, GB/2423.13-2002 temperature impact test, GJB360B-2009 temperature impact test, GB 2423.1-2 008/IEC6008- 2-1-2007 electrical and electronic products environmental test Part 2: Test methods, Test A: Low temperature, GBT 2423.2-2008 electrical and electronic products environmental test Part 2: Test methods, Test B: High temperature, GBT 10589-2008 low-temperature test chamber technical conditions, GBT 11158-2008 high-temperature test chamber technical conditions
1. Confirm that the installation environment of the equipment meets the requirements (Ambient Temperature 0~ 35 ℃, relative humidity ≤ 85%, and keep a sufficient distance between the box and the wall). 2. Connect the required power supply (AC380V three-phase + protective grounding), and configure an independently controlled air or power switch. 3. If you buy a Water-Cooled model, you need to connect a chilled down water tower (Flow rate 8 tons/hour). 4. Perform human-machine interaction through 7-inch Touchscreen controller. 5. Edit or select the test program (up to 120 groups) as needed, and set parameters such as Temperature range, time, and number of cycles. 6. Place the test sample on a stainless steel rack (load-bearing ≤ 20kg/piece) and close the box door. 7. Select the starting mode (high temperature/low temperature/normal temperature start) and start the test program. 8. During the operation, the temperature, Linear dispersion, program segment, remaining time and other information can be monitored in real time through the screen. 9. After the test is completed, the device can be automatically stopped or manually stopped according to the setting. 10. Export test data through U disk or Communication interface.
* Equipment Operating Environment Temperature should be kept within 0 ℃~ + 35 ℃, relative humidity ≤ 85%. Ambient Temperature should be kept within 0 ℃~ + 45 ℃; if Ambient Temperature is lower than 0 ℃ and the equipment is stopped for a long time, the water retained in the equipment must be drained clean to prevent the pipeline from freezing. * When installing, the distance between the left and right side panels of the test chamber and the rear side panels shall not be less than 500mm, and the distance between the front and the wall shall not be less than 1000mm. Make sure that the power supply meets the requirements (AC380V three-phase, voltage allowable fluctuation range +/- 10%), and protect the grounding (grounding resistance is less than 4Ω). Users need to configure the equipment with independent air or power switches of corresponding capacity on site. When the equipment is running, please pay attention to the size of the entrance or passage, elevator, etc., to ensure that the equipment can be handled and installed smoothly. * Do not exceed the nominal load-bearing capacity of the shelf (each piece is less than or equal to 20kg). * When the equipment fails, the controller will display the fault screen and alarm, please check and deal with it according to the prompt information.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.