$ 16291.00Ex works
The double-channel conversion structure is used to achieve fast temperature impact, the conversion time is no more than 10 seconds, the temperature control accuracy is +/- 0.5 ℃, and 120 sets of program storage and USB data export functions are supported.
Category : Thermal Shock Test Chamber
Brand : HUDA
SKU : NB041402
Service :The delivery and service By NBCLab.Com
HUDA QD-E703-100K55 Thermal Shock Test Chamber is a testing equipment required by the metal, plastic, rubber, electronics and other material industries. It tests the impact of rapid changes in high and low temperatures on products in a way that the items to be tested do not move. The equipment adopts computer control, is equipped with self-developed software, and has a good operation interface. It supports parameter setting, program keeping and automatic operation functions.
1. Adopt imported LED digital display microcomputer integrated controller, touch input is easy to operate 2. The maximum program capacity is 120 groups, and it supports connection and use between programs 3. With USB storage function, it can back up parameters and record operation Linear dispersion 4. Equipped with RS-232/RS-485 communication interface to support remote monitoring 5. Using environmentally friendly refrigerant R404A/R23 6. Multiple safety protection devices, including Compressor protection, over-temperature protection and electrical protection 7. With reservation start/stop function and power-off program memory function 8. Imported core components, including French Taikang Compressor and Taiwan Youyi controller
The device moves the test basket between the high temperature groove and the low temperature groove through the pneumatic cylinder actuation mechanism to realize the Thermal Shock test. Built-in high temperature zone, low temperature zone air valve, loop air duct and stainless steel loop fan, the test product is sent to the heat and cooling space by switching the basket, so as to achieve rapid temperature exchange and uniform indoor temperature distribution.
Electronic products, automotive parts, high-tech products, metal materials, plastic materials, rubber materials
MIL STD, GB, JIS, JEDEC, IEC, GB/T 2423.1 - 1989 Low Temperature Test Method, GB/T 2423.2 - 1989 High Temperature Test Method, GB/T 2423.22 - 1989 Temperature Change Test, GJB150.5-86 Temperature Impact Test, GJB360.7-87 Temperature Impact Test, GJB367.2-87 405 Temperature Impact Test, Japanese Industrial Standard JIS, Industrial Standard JESD22, Communication Industry Standard GR-1221-CORE, GB T 10589-2008 Low Temperature Test Chamber Technical Conditions, GB T 11158-2008 High-temperature test chamber Technical Conditions
1. Set the time and parameters through the touchscreen 2. Select the program mode and start running 3. Equipment automatically performs cooling, heating, and temperature switching 4. Real-time monitoring of temperature Linear dispersion and data 5. Data can be exported via USB after the test is completed
Equipment environment temperature should be kept within 0 ℃~ 35 ℃, relative humidity ≤ 85% When the environment temperature is lower than 0 ℃, the water left in the equipment should be drained clean • The distance between the left and right sides of the test chamber and the rear side panels shall not be less than 500mm, and the front distance shall not be less than 1000mm • Confirm equipment location size and channel requirements before installation • Power supply voltage is 3-phase 5-wire 380V, need to ensure stable power supply • Regularly check the refrigeration system and safety protection devices
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.