$ 6861.00Ex works
Adopt cascade refrigeration system, temperature control accuracy +/- 1 ℃, humidity control accuracy +/- 2% RH, with 1000 segment program control ability, support RS-232 communication interface and automatic calculation function.
Category : Temperature Humidity Chamber
Brand : kesionots
SKU : NB041481
Service :The delivery and service By NBCLab.Com
KESIONOTS KS-HW225L (150 ℃) humidity chamber is a testing equipment that can accurately simulate complex natural environments such as low temperature, high temperature, high temperature and high humidity, low temperature and low humidity. It is suitable for product Reliability testing in plastics, electronics, food, clothing, vehicles, metals, chemicals, building materials and other industries. It has advanced temperature and humidity control systems and various safety protection functions.
1. With automatic calculation function 2. With fixed temperature and slope setting function 3. With hold, skip, standby and two sets of time signal output functions 4. With cold/heat engine start-up selection function 5. With temperature display value correction function 6. With temperature sensor calibration function 7. With Compressor automatic stop function 8. With reservation start/stop function 9. With power failure memory function, program data can be maintained for 6 months after power failure 10. Powerful network monitoring and control capabilities 11. With 1000 program segments, each segment can cycle through 999 steps 12. With fault alarm and cause, processing prompt function 13. With upper and lower limit temperature protection function 14. With self-diagnostic function 15. With PC link function
The test chamber adopts a cascade refrigeration system, which is composed of a high temperature part and a low temperature part. The refrigeration cycle is realized by connecting the evaporative condenser. The heating system is composed of high-power resistance wire, which can heat up quickly. The temperature and humidity system is divided into two subsystems of humidification and dehumidification. The humidification adopts the steam humidification method, and the dehumidification mainly adopts the mechanical refrigeration and dehumidification method. The Sensor system adopts platinum resistance and Thermocouple measurement temperature, and the solid-state electronic Sensor measurement humidity.
Reliability testing of products in plastics, electronics, food, clothing, vehicles, metals, chemicals, building materials and other industries
GB5170、2、3、5、6-95、GB2423.1-89 (IEC68-2-1)、GB2423.2-89 (IEC68-2-2)、GB/T2423.3-93(IEC68-2-3)、GB/T423.4-93(IEC68-2-30)、GJB150.3(ML-STD0-810D)、GJB150.4(MIL-STD-810D)、GJB150.9-1986、GJB4.6-1983、GJB367.2-1987、GJB4.5-1983、GB10592-93、GB10586-93、GJB360.8-87(MIL-STD-202f)、GB/T5170.2-96、GB/T5170.5-96、YD1268-2003、UL1642、UN38.3、GB/T18287-2000、GB8897.4-2002、UL2054-2005、SJ/T 11169-1998、GB8897.4-2002、SJ/T11170-1998、YD1268.2-2003、QC/T743-2006、QC/T744-2006
1. Turn on the power supply and check the grounding resistance ≤ 4Ω. 2. Manually add water to 20L capacity 3. Set test program parameters 4. Select the start-up mode (cold/heat) 5. Start the test program 6. Monitor test process Linear dispersion 7. Turn off the device after the test is completed
• Power supply voltage must be kept within 220V AC (+/- 10%) range • Must use air switch (40A), no knife switch or power socket • The resistivity of humidification water needs to be ≥ 500MΩ • The weight of the product rack shall not exceed 20 kg • The equipment has multiple safety devices such as air burn prevention, high voltage protection, overheating protection, and overcurrent protection • The start-stop function can be locked during operation to prevent accidental touch from affecting the experimental results • Operating temperature range -70~ 150 ℃, Humidity range 20~ 98% R.H • Regularly check whether the equipment safety devices are normal
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.