Overview
KESIONOTS KS-OTS-LR80LC Thermal Shock Test Chamber is used to test the tolerance of material structures or composites in instantaneous extremely high temperature and extremely low temperature continuous environments, and to conduct rapid tests through chemical changes or physical damage caused by thermal expansion and contraction. Suitable for metals, plastics, rubber, electronics and other materials, providing basis for product improvement.
Features
1. Adopt a three-box equipment structure, including a high-temperature zone, a low-temperature zone, and a test zone, to test the static placement of the product.
2. Touch graphic control operation interface, easy to operate.
3. Impact mode uses wind path switching to introduce temperature into the test area for Thermal Shock test.
4. Temperature recovery time ≤ 5min, temperature conversion time ≤ 10s, temperature control accuracy +/- 0.5 ℃.
5. The liner material is stainless steel, and the outer box material is SUS304 #stainless steel plate matte surface treatment or cold tie steel plate nanotechnology advanced baking paint.
6. The thermal insulation material is made of high temperature and high density chlorine-based formic acid vinegar foam insulator material.
7. The system adopts P.I.D + S.S.R + microcomputer equilibrate temperature control system.
8. The chilling down system can be selected from semi-hermetic double-stage Compressor (water-cooled) or fully hermetic double-stage Compressor (air-cooled).
9. Equipped with a variety of protection devices, including no fuse switch, Compressor high and low voltage protection switch, refrigerant high voltage protection switch, fault warning system, electronic alarm.
10. Provide accessories such as viewing windows, upper and lower adjustable compartments, power-on measuring holes, casters, and horizontal brackets.
Principle
The equipment adopts three-box structure, the test product is placed in the test area, and the temperature of the high temperature zone or the low temperature zone is introduced into the test area for Impact test through the air passage switching mode, so as to realize the simulation of the alternating hot and cold environment.
Applications
Metal, plastic, rubber, electronics
Steps
1. Place the test product in the test area.
2. Set the temperature parameters of the high temperature zone and the low temperature zone through the touch graphic operation interface.
3. Start the equipment, and the air circuit switching system will introduce the temperature into the test area for Impact testing.
4. Monitor the temperature recovery and conversion process to ensure accuracy.
5. After the test is completed, remove the product and analyze the results.
Notice
• Power supply voltage is AC 380V 50HZ/60HZ.
• The weight of the equipment is 900kg, which needs to be moved and fixed with casters and horizontal brackets.
• Check protective devices such as fuse-free switches and Compressor protection switches before operation.
• Avoid use in extreme environment conditions to prevent affecting test accuracy.
• Maintain the chilled down system and insulation materials regularly to ensure long-term stable operation of the equipment.