Please contact customer service
With 0~ 100 ℃ automatic temperature compensation and range automatic switching function, measurement error +/- 2.0% F.S., IP65 protective shell and photoelectric isolation technology, strong anti-interference ability, support Multi-ParaMeter display and relay control.
Category : Conductivity Meter
Brand : NOBOTECH
SKU : NB041735
Service :The delivery and service By NBCLab.Com
CD-5000 is an industrial online intelligent conductivity /TDS meter independently developed by NOBOTECH brand. The instrument adopts high-performance CPU chip and SMT patch technology, which can accurately measure conductivity /TDS and temperature, and has the functions of temperature compensation and range automatic conversion. It has the characteristics of high accuracy and good Repeatability. It can be widely used in continuous monitoring of conductivity value/TDS in solutions such as thermal power, chemical fertilizers, metallurgy, environmental protection, biochemistry, food and tap water.
1. Adopt new waterproof, dustproof and moisture-proof good shell (IP65) process, surface painting, Appearance generous high-end. 2. The product has High Accuracy, High Stability performance and strong anti-interference ability. 3. Large screen point matrix LCD display, support Chinese and English menus (switch between each other), Multi-Parameter display at the same time. 4. Can carry out conductivity /TDS measurement, upper and lower limit control, range automatic switching, improve measurement resolution. 5. The screen displays the alarm status and can be accompanied by the switch ON/OFF signal output at the same time. 6. With automatic temperature compensation function, the range is 0~ 100 ℃. 7. Powerful communication function: Both RS485 and 4-20mA interfaces use photoelectric isolation technology. RS-485 Communication interface is compatible with MODBUS part protocol (optional), and can be converted to RS-232. The value corresponding to the current output of 4~ 20 mA can be set arbitrarily. Support the company's internal Internet of Things and customer-customized Internet of Things. 8. The hysteresis can be set arbitrarily to avoid frequent operation of the switch relay. 9. Can be customized for a large range (within 0-300 mS · cm-1). 10. Equipped with watchdog function to ensure that the meter does not crash. 11. The core components are all from foreign brands. 12. Factory settings can be restored. Power failure protection > 10 years.
CD-5000 online conductivity /TDS uses high-performance CPU chip and SMT patch technology to measure the conductivity /TDS and temperature of the solution. The meter has a built-in NTC10k temperature sensor, which is based on 25 ° C and automatically performs temperature compensation in the range of 0~ 100 ° C. The temperature compensation coefficient can be corrected. During the measurement process, the meter can automatically switch the range according to the signal size, and the measurement result can be displayed through a large-screen LCD display. At the same time, it can output control signals or data through relays, 4-20mA current or RS485 Communication interface.
Thermal power, chemical fertilizers, metallurgy, environmental protection, biochemistry, food, tap water
1. Install the instrument and ensure that the opening size is 92 x 92mm. 2. Select the appropriate Electrode (such as 0.01, 0.1, 1.0 or 10.0 Electrode) according to the measurement requirements and connect. 3. Connect the Power Supply (220VAC, DC24V or DC12V). 4. After the instrument is powered on, the parameters are set through the large-screen LCD menu, such as range, temperature compensation coefficient, relay control point, hysteresis, 4-20mA output corresponding value, etc. 5. Immerse Electrode in the solution to be tested. 6. The instrument automatically performs measurement, temperature compensation and range conversion, and displays the results. 7. Set the automatic cleaning function (cleaning time and interval) as required. 8. Connect external devices via relay output, 4-20mA current or RS485 Communication interface.
* Please ensure that the working Ambient Temperature is 0~ 60 ℃, relative humidity ≤ 90%. * 4~ 20mA current output, the load should be < 750Ω. * Relay load is 3A-220VAC, do not overload when using. The meter has a power-down protection function, but do not power off frequently unless necessary. * When making large-scale measurements (such as exceeding the standard range), it is necessary to customize to the manufacturer in advance. * When installing, please note that the Protection Rating of the meter is IP65, but you should still avoid long-term immersion or extreme environments. When using RS485 for communication, it is necessary to confirm protocol compatibility (MODBUS partial protocol). * If you need to restore the factory settings, please follow the instructions.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.