Overview
The LC-LGH-300 is a gas Chromatograph gas source that replaces high voltage cylinders, making analysis more convenient and safe. It is equipped with a cold drying system to process the generated gas through condensation adsorption technology without frequent replacement of discoloration silica gel.
Features
1.7-Inch color display screen, real-time monitoring of cold drying temperature, output pressure, output Flow rate, and display of cold drying, electrolysis, water level status and online leak detection.
2. Adopt precision pressure sensor digital display pressure, equipped with atmospheric pressure starting device to prolong Compressor life.
3. With automatic water replenishment function, no manual operation is required.
4. The output gas temperature is low, the dew point is better than -70 ℃, and the water content is low.
5. With the boot time and cumulative Working time memory function.
6. Adopt low temperature cold drying, pressure swing adsorption and water removal technology to reduce the workload of replacing discoloration silica gel.
7. Purification tube is made of high strength PC material, mold injection molding, Reagent resistance, high temperature resistance, Impact resistance.
8. Simple operation, gas production can be started, and the output pressure is stable, which can meet the needs of multiple Chromatographs.
9. Add potassium hydroxide solution for the first time, and then add distillation water to use continuously.
10. Equipped with hydrogen leakage protection function, automatically cut off the power supply and alarm when leaking.
11. With water protection function, protect the Gas Chromatograph.
12. Equipped with a non-return liquid check valve to avoid the phenomenon of return liquid.
13. Replace the traditional high voltage hydrogen cylinder to achieve laboratory instrumentation.
14. Cold drying, water replenishment, electrolysis, oxygen integration fully automatic operation, to achieve 24 hours unattended.
Principle
The hydrogen generator uses electrolysis water to generate hydrogen and is equipped with a cold drying system. The generated hydrogen is treated by low temperature cold drying and pressure swing adsorption technology to remove water, so that the gas is condensed and adsorbed, thereby reducing the water content and eliminating the need for frequent replacement of discoloration silica gel.
Notice
• Ambient Temperature should be 5 ℃~ 40 ℃, Humidity ≤ 85%.
• Must use distillation or deionized water.
• Note that the specification of the gas pipe is 3mm in outer diameter, and the connector is M8 × 1 nut.