Temperature range -40~ 150 ℃, humidity control 20~ 98% RH, using French imported Compressor and environmentally friendly refrigerant, equipped with 7-inch touch screen controller imported from Korea, support 100 sets of programs and 5000 segments editing, with power-off memory and Linear dispersion display function.
Category : Temperature Humidity Chamber
Brand : JiangKai
SKU : NB042198
Service :The delivery and service By NBCLab.Com
JIANGKAI NBC-HW-150 (-40 ℃) Constant Temperature and Humidity Chamber is a test equipment used to simulate high temperature, low temperature and constant, alternating heat and humidity environment. It mainly conducts environmental adaptability tests on parts and materials of electronic, electrical, household appliances, automobiles and other products, and inspects its performance indicators.
1. Adopt fully dense Compressor imported from France, environmentally friendly refrigerant R404A, equipped with fin condenser and evaporator to form an efficient and stable refrigeration system. 2. The core control system is the original Korean imported TEMI880 all Chinese LCD 7-inch touch screen, the operation interface is friendly, and it supports both temperature and humidity program control and Linear dispersion display. 3. The controller has high accuracy, temperature +/- 0.1 ℃, humidity +/- 1% RH, with multiple sets of PID control, automatic calculation and automatic correction of wet and dry bulbs. 4. Large program capacity, supports up to 100 sets of programs, 5000 segments (SEGMENTS), each segment can be set up to 530 hours and 59 minutes, and has power-off memory and connection execution functions. 5. The inner box is made of stainless steel plate (SUS #304), the insulation layer is rigid polyurethane foam and Glass cotton, and the box door is equipped with three-layer vacuum observation window, which has good insulation and observation performance.
The equipment works together with the refrigeration system (Compressor, condenser, evaporator, etc.) and the heating system (heating rod) to achieve precise control of the temperature in the test chamber. The humidity control is regulated by the humidification system (humidifier tube) and the dehumidification system (auxiliary to the refrigeration system). The core Korean TEMI880 controller receives signals from the PT100 temperature sensor and humidity sensor. After PID operation, it outputs control instructions to the refrigeration, heating, humidification and other executive components, thereby creating and maintaining the set temperature and humidity environment in the box to simulate various climate conditions that the product may encounter in actual use.
Reliability testing of electronic, electrical, household appliances, automobiles and other product parts and materials
GB5170.2, GB5170.3, GB5170.5, GB5170.6-95 "electrical and electronic products environmental test equipment basic parameters method of verification", GB 2423.1-89 (IEC68-2-1) test A: low temperature test method, GB 2423.2-89 (IEC68-2-2) test B: high temperature test method, GB/T 2423.3-93 (IEC68-2-3) test Ca: constant humidity cabinet test method, GB/T 2423.4-93 (IEC68-2-30) test Da: alternating humidity cabinet test method
1. Enter the program editing screen through the touch screen, and directly enter or edit the required temperature and humidity parameters and time. 2. Select or set the program group to be executed, and confirm parameters such as the number of cycles. 3. Put the sample to be tested into the test chamber and close the box door. 4. Start the set test program on the main interface of the controller. 5. During the test, the set value, actual value and program execution Linear dispersion can be observed in real time through the screen. 6. After the program runs, the device stops or stands by according to the setting, and the sample is taken out.
* The equipment should be installed in a well-ventilated place without direct sunlight and away from heat sources. When placing the sample, ensure that the sample placement does not affect the air circulation in the box, and the total load should not exceed the specified limit. When the equipment is running, do not open the door frequently to avoid affecting the stability of temperature and humidity and the lifespan of the equipment. Regularly check and clean the condenser, humidifying water tray and other components to ensure the normal operation of the equipment. * If the device fails, the screen will display an alarm message, please contact a professional according to the prompts, and do not disassemble and repair it yourself. When not in use for a long time, the humidified water should be drained and powered on regularly.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.