$ 19339.00Ex works
The three-box structure is used to realize the static test sample, the heat storage and cold storage design is combined with the forced air circuit switching, the temperature conversion time is less than or equal to 10 seconds, the temperature control accuracy is +/- 0.5 ℃, and the impact conditions are independently set in the high temperature and low temperature regions.
Category : Thermal Shock Test Chamber
Brand : Apkjgroup
SKU : NB042363
Service :The delivery and service By NBCLab.Com
APKJ 3AP-CJ-50C Thermal Shock Test Chamber is a device used to test the resistance performance of materials in extremely high temperature and extremely low temperature continuous environments. It adopts a three-box structure, which is divided into high temperature zone, low temperature zone and test zone. It is introduced into the test zone by strong cooling hot air circuit switching to realize rapid temperature Impact testing. It is widely used in Electronics, Auto Parts, Metal Chemical Materials and other industries.
1. Three-box equipment structure, the test sample is completely still in the test area, using a unique heat storage and cold storage structure, and the strong cooling hot air circuit switching mode is introduced into the test area. 2. It can independently set three different conditions of high temperature, low temperature and Thermal Shock. When executing Thermal Shock, you can choose 2 or 3 boxes, and it has the function of high and low temperature testing machine to support normal temperature Impact. 3. The use of rigid polyurethane foam insulation material has the advantages of energy saving, shock and compression resistance, ultra-low temperature thermal conductivity, weather resistance and heat preservation, insulation and sound insulation, waterproof and moisture-proof. 4. Equipped with self-developed 7-inch super touch AP-950 programmable temperature controller, it supports touch screen input, temperature alarm display and real-time temperature Linear dispersion display. 5. Multiple safety protection devices, including Compressor overpressure, overheating, overcurrent protection, laboratory temperature is too high and too low protection, total power supply phase sequence and phase loss protection.
The Thermal Shock Test Chamber adopts the structure of heat storage and cold storage, and quickly introduces the temperature in the high temperature or low temperature region into the test area for Impact through the switching mode of strong cooling hot air circuit. The test sample is still in the test area, and the chemical or physical changes caused by thermal expansion and contraction are tested in an instant through an extremely high temperature and extremely low temperature continuous environment. The temperature recovery time is short, and the conversion temperature only needs ≤ 10 seconds.
Electronics components plastic and other industries, automation parts, communication components, auto parts, metals, chemical materials, plastics and other industries, national defense industry, aerospace, military industry, BGA, PCB base wrench, electronic chip IC, semiconductor ceramic and polymer materials
GB/T2423.1-2008、GB/T2423.2-2008、GB/T2423.22-2012、GJB150.5-86、GJB360.7-87、GJB367.2-87、QC/T17-92、EIA364-32、IEC68-2-14
1. Configure the corresponding capacity air or power switch to ensure an independent supply of AC 380V power. 2. Connect the power cord, check the voltage, frequency and grounding resistance of the protection wire. 3. Set the high and low temperature Impact temperature point and program through the touch screen controller. 4. Place the sample in the test area and start the device to perform the Thermal Shock test. 5. Monitor real-time temperature Linear dispersion and alarm information, and automatically stop after the test is completed.
• Installation site requires an independent AC380V power supply, voltage range AC (380 +/- 10) V, frequency (50 +/- 0.5) Hz. • The grounding resistance of the protective wire should be less than 4Ω to ensure safety. When the equipment is not working, the environment temperature should be kept at 0 ℃~ + 45 ℃, and when it is lower than 0 ℃, water should be discharged to prevent the pipeline from freezing. • External environment items such as ventilated pipes, power supplies, gas sources, and water sources need to be installed by users. • Pay attention to the status of the safety protection device during operation to avoid overheating or abnormal power supply.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.