Overview
CHENCHI WDK-01 (3000N) tensile strength testing machine is designed for universities and research institutes. It has beautiful appearance, convenient operation and stable and reliable performance. It has a wide range of adaptations and complete functions. It can detect the tensile strength, tensile strength, elongation, fracture length, tensile energy absorption, tensile index, tensile energy absorption index, fracture time and other parameters of the paper. The system controls the servo motor to drive the precision lead screw to realize the sample stretching through the CHENCHI controller. It has the characteristics of no pollution, low noise, high efficiency and wide speed regulation range.
Features
1. 高精度和可重复性:采用高精度测力传感器,准确度<0.5%,优于ISO标准±1%;进口步进电机和细分技术确保定位准确、速度响应快。
2. 寿命长:精密滚珠丝杆传动,合理设计和机加工工艺保证耐用性,支持升级。
3. 适应范围广,功能齐全:多种量程可选,适用于所有纸和纸板,检测参数全面。
4. 使用方便,易于操作:高速ARM处理器和5吋触摸液晶屏,自动化程度高,支持数据统计和打印输出。
5. 测试快速有效:可预设试验速度、样品宽度和克重,内置多单位转换功能。
6. 便于计量校准:测力传感器位于上方,可轻松挂载标准砝码进行校准。
Principle
The system controls the speed control system to drive the servo motor to rotate through the CHENCHI controller. After deceleration by the deceleration system, the beam is driven up or down by the precision lead screw pair to complete the tensile mechanical properties test of the sample.
Applications
Universities and research institutes
Standards
GB 8808、GB/T 1040.1、GB/T 1040.2、GB/T 1040.3、GB/T 1040.4、GB/T 1040.5、GB/T 4850、GB/T 12914、GB/T 17200、GB/T 16578.1、GB/T 7122、GB/T 2790、GB/T 2791、GB/T 2792、GB/T 17590、ASTM E4、ASTM D882、ASTM D1938、ASTM D3330、ASTM F88、ASTM F904、JIS P8113、QB/T 2358、QB/T 1130
Steps
1. Connect the power supply and start the device.
2. Set test parameters such as speed, sample width, and gram weight through a 5-inch touch LCD screen.
3. Install the specimen onto the standard fixture.
4. Start the test, the system automatically completes the tensile test and collects data.
5. View or print test results, optionally connect a computer to display Linear dispersion and save data.
Notice
• Check the power connection before use to ensure compliance with 220V 50Hz requirements.
• Pay attention to whether the overload protection, leakage protection and limit protection functions are normal during operation.
• Measurement calibrated regularly, using standard weight hanging on the upper chuck to verify accuracy.
• Avoid overloading to prolong the life of the equipment.
• Keep equipment clean to prevent dust from affecting performance.