Overview
TAISITE WGL-85D electric air blowing dryer is a vertical drying equipment with PID intelligent temperature control. Temperature range RT + 10~ 300 ℃, temperature resolution 0.1 ℃, Temperature Fluctuation +/- 1.0 ℃, temperature distribution accuracy +/- 2.0%. Inner chamber dimensions 450 * 450 * 450mm, volume 85L, suitable for drying, heat treatment and other applications in laboratories, industries and other fields.
Features
1. Adopt PID intelligent temperature control to quickly reach the set value and operate stably
2. Convenient operation, support fixed-value operation, regular operation, automatic stop, appointment start
3. Dedicated function keys to achieve temperature setting
4. Auxiliary menu, realize over-rise alarm, deviation correction, menu lock
5. Vertical double air duct air sleeve, turbine centrifugal forced air supply, more uniform temperature
6. Speed-regulating fans enable the use of lightweight, fine, powdery samples
7. Forced fan cooling intake structure, so that the fan maximum Operating temperature < 50 ℃ long life operation
8. Insulation door technology, when used at high temperatures, the outer door temperature is low and does not deform
9. Pull-out tray Shelf design, easy and fast to take samples
10. Self-developed digital display overlift preventer makes temperature control more accurate, stable, safe and reliable (optional)
11. Over-rise alarm, menu lock, over-rise prevention (optional)
Principle
The equipment adopts double air duct forced convection method, which is heated by stainless steel heating pipe, and the turbine centrifugal forced air supply system makes the hot air circulate evenly in the studio. The PID intelligent temperature control system precisely adjusts the temperature to ensure that the temperature is stable at the set value. The speed regulating fan can adjust the wind speed according to the characteristics of the sample, and the forced heat dissipation structure ensures the long-term stable operation of the fan.
Applications
Laboratory, industry, scientific research, education, medicine, chemical industry, electronics, materials
Standards
Patent number: ZL201630486063.2, ZL201320059701.3, ZL201921000740.X, ZL201921001075.6, ZL201920940767.0
Steps
1. Turn on the power supply and set the desired temperature
2. Select the operation mode: fixed-value operation or regular operation
3. Set the timing time (if necessary)
4. Start the device and wait for the temperature to reach the set value
5. Put in the sample for drying
6. The device automatically stops or manually shuts down after processing is completed
Notice
• Power supply voltage meets AC220V requirements before use
• Do not overload, Load per shelf 15kg
• Avoid direct contact with the outer door during high temperature operation
• Regularly check the operating status of the fan
• Keep the exhaust port unobstructed
Do not move or tilt the device while it is running
• The optional function needs to be confirmed before use.