Overview
Kenton DHG-9130C + independent temperature limiter thermostatic drying oven is a high-performance laboratory equipment, using microcomputer PID digital display temperature control instrument, Temperature range RT + 5~ 250 ℃, temperature resolution +/- 0.5 ℃, Temperature Fluctuation +/- 1%, studio volume 130L, suitable for precise temperature control drying, heating and other processes.
Features
1. Microcomputer PID digital display temperature controller, accurate and stable temperature control
2. Independent temperature controller, standard safety protection
3. The studio is made of mirror stainless steel, which is corrosion-resistant and easy to clean
4. Equipped with high temperature bearings and cooling fans to extend service life
5. Three-sided + front air circulation duct, good temperature uniformity
6. Silicone door sealing and stable cotton heat insulation material, excellent thermal insulation performance
7. Timing function 0~ 999min, flexible and convenient operation
Principle
The equipment precisely sets and adjusts the temperature through a microcomputer PID temperature controller. The electric heating tube (201 stainless steel) generates heat. Combined with the three-sided + front air outlet circulating air duct system, the hot air is evenly distributed in the studio (mirror stainless steel) to achieve thermostatic drying. The independent temperature limiting controller provides additional temperature protection to ensure safe operation.
Applications
Laboratories, Research institutes, Industrial production, Material drying, Sample Handling
Steps
1. Connect single-phase AC220V/50Hz power supply
2. Turn on the power switch of the device
3. Set the required temperature through the microcomputer PID thermostat (RT + 5~ 250 ° C)
4. If you need timing, set the Timer Range 0~ 999min
5. Start the heating, and the equipment automatically controls the temperature to the set value
6. Turn off the power after use
Notice
The working environment humidity should be less than 70% RH
• Power supply voltage is single phase AC220V/50Hz
• Regularly check the status of door sealing silicone and heat insulation material
• Avoid overloading and ventilate the equipment
• Use a soft cloth when cleaning to avoid damage to the mirror stainless steel studio