Overview
HUITAI HHG-9078A + high temperature blast drying oven is a drying equipment equipped with independent temperature limiting controller, suitable for drying, baking, wax, sterilizing and other operations in industrial and mining enterprises, laboratories, R & D institutions and other places.
Features
1. The studio is made of high-quality steel or stainless steel.
2. Microcomputer intelligent temperature controller, with dual digital display, timing, power suppression and self-tuning functions for setting and measuring temperature, accurate and reliable temperature control.
3. The hot air circulation system is composed of low-noise fans and air ducts to effectively ensure uniform temperature in the working room.
4. Independent temperature limit alarm system, which automatically interrupts when the temperature exceeds the limit, ensuring the safe operation of the experiment without accidents.
5. Optional with RS485 interface, which can be connected to a recorder and a computer to record changes in temperature parameters.
Principle
The equipment precisely sets and controls the heating temperature through a microcomputer temperature controller, and uses a low-noise fan to drive the hot air to circulate inside the studio to ensure uniform temperature distribution. When the temperature exceeds the safety threshold set by the independent temperature limiting controller, the system will automatically interrupt the heating to ensure safety.
Applications
Industrial and mining enterprises, laboratories, R & D institutions
Steps
1. Place the device in a stable, well-ventilated position and connect the power supply.
2. Open the box door, put in the items to be processed, and be careful not to block the air duct.
3. Close the box door and set the required temperature and timing time through the microcomputer thermostat.
4. Start the equipment and start the heating and blast circulation.
5. During the working process, the real-time temperature can be observed through the display screen.
6. After the set time is reached or the work is completed, the equipment stops heating and the items are taken out after the temperature drops naturally.
Notice
Equipment should be placed in a well-ventilated, flammable and explosive-free environment.
• Do not block the goods in and out of the trend, so as not to affect the hot air circulation and temperature uniformity.
• The items placed should be resistant to high temperature and should not be placed too tightly.
• During operation and just at the end, the external and internal temperatures of the box are high, so do not touch it to prevent burns.
• When using an independent temperature limit alarm system, please set the temperature limit value correctly.
• Be sure to disconnect the power supply before cleaning, until the box is completely chilled down.
• If the RS485 interface is selected for data logging, please make sure the connection is correct.
• Non-professional personnel should not disassemble or repair the equipment.