Overview
HUITAI GDT-150C high and low temperature alternating climate box is a professional temperature and humidity testing equipment, providing a wide Temperature range of -60~ 150 ℃ and a relative humidity control of 20%~ 98% (model with H). It adopts a 7-inch TFT color LCD display, supports setting mode and program mode operation, and has data recording and remote communication functions.
Features
1. Temperature range -60~ 150 ℃
2.7-Inch TFT color LCD display, can be connected to RMCS control software
3. Support fixed value mode and program mode, preset 100 groups of 100 steps 999 cycles
4. Adopt BTC equilibrate temperature control + DCC intelligent cooling capacity control + DEC intelligent electrical control
5. RAM data logging function with battery protection, the maximum recording time is 350 days
6. Support RS-485 and Ethernet communication interfaces, and provide host computer operating software
7. France Taikang fully enclosed Compressor refrigeration system
8. Heated observation window with LED interior lighting
9. Multiple safety protection systems include over-temperature protection, Compressor protection, leakage protection, etc
Principle
The equipment adopts BTC equilibrate temperature control mode combined with DCC intelligent cooling capacity control and DEC intelligent electrical control, and realizes temperature regulation through mechanical single-stage compression or cascade Compressor refrigeration system. Humidity control adopts wet-bulb temperature meter method, and temperature and humidity sensing is carried out through A-class PT100 armored Thermocouple. The refrigeration system adopts French Taikang fully enclosed Compressor, combined with slope evaporator and water-cooled condenser, to ensure stable operation through nitrogen flushing welding process and shock absorption design.
Applications
Electronic product testing, material research, automotive parts testing, aerospace parts testing, military product testing, environmental adaptability testing, quality control testing
Standards
GB/T5170.2-2008、GB/2423.1-2008、GB/2423.2-2008、GB/T5170.5-2008、GJB150.3、GJB150.4、GB2423.3-93、GB2423.4-93
Steps
1. Connect the power supply and check whether the voltage meets the 220V requirements
2. Set operating parameters through a 7-inch TFT color LCD display
3. Select setting mode or program mode to run
4. Start the device and monitor the operating status
5. Export test data via USB interface or Ethernet
6. Turn off the device according to the procedures after completing the test
Notice
• Check the power supply phase sequence and phase loss protection function before use
• Ensure that the water supply system is normal and avoid dry burning of the humidifier pipe
• Regularly check the refrigeration system pressure and protection devices
• Do not frequently switch on and off the observation window when the equipment is running
• Keep equipment well ventilated
• Place test samples strictly according to load capacity requirements
• Regular equipment calibration and maintenance
• Pay attention to the functional status of equipment leakage protection and lightning protection