HUITAI GDT-225C high and low temperature alternating climate box is a professional temperature and humidity testing equipment, providing a wide Temperature range of -60~ 150 ℃ and 20%~ 98% relative humidity control. Equipped with 7-inch TFT color LCD display and intelligent control system, it supports setting mode, program mode and remote communication function, suitable for various environment Reliability test.
Features
1. Adopt BTC equilibrate temperature control + DCC intelligent cooling capacity control + DEC intelligent electrical control
2.7-Inch TFT color LCD display with support for RMCS control software
3. Support fixed value mode and program mode, preset 100 groups of 100 steps 999 cycles
4. With battery protection RAM, the maximum recording time is 350 days
5. France Taikang fully enclosed Compressor, high efficiency refrigeration system
6. Multiple safety protection systems, including Compressor protection, humidification system protection and test chamber protection
7. Support RS-485 and Ethernet communication, compatible with a variety of operating systems
Principle
The equipment adopts a mechanical compression refrigeration system, and realizes accurate temperature and humidity regulation through BTC equilibrate temperature control mode combined with DCC intelligent cooling capacity control and DEC intelligent electrical control. The temperature detection adopts A-class PT100 armored Thermocouple, and the humidity detection adopts wet-bulb thermometer method. The refrigeration system adopts high-efficiency slope evaporator and SWEP plate cold and heat exchanger to ensure fast and stable temperature change.
Applications
Electronic product testing, automotive parts testing, material research, aerospace, military product testing, communication equipment testing, medical apparatus testing
Standards
GB/T5170.2-2008、GB/2423.1-2008、GB/2423.2-2008、GB/T5170.5-2008、GJB150.3、GJB150.4、GB2423.3-93、GB2423.4-93
Steps
1. Connect power 380V/32A
2. Set test parameters through a 7-inch TFT color LCD display
3. Select setting mode or program mode to run
4. Monitor equipment operating status and data logging
5. Export test data via USB or Ethernet interface
6. Turn off the device according to the procedures after the test is completed
Notice
Equipment must be used under standard load conditions
• Pay attention to the phase sequence and phase loss protection of the power supply
• Regularly check the water supply of the humidification system
• Avoid Compressor overpressure operation
• Pay attention to equipment heat dissipation and ventilating requirements
• Confirm that all safety protection functions are normal before testing
• Execute test procedures strictly according to the operation manual